US · guidance
CMS SOM App. H, Tag V403
§494.60 (b) Standard: Equipment maintenance
The dialysis facility must implement and
maintain a program to ensure that all equipment (including emergency equipment, dialysis
machines and equipment, and the water treatment system) are maintained and operated in
accordance with the manufacturer’s recommendations.
Interpretive Guidance §494.60 (b)
All equipment used at the dialysis facility and dialysis equipment used by home dialysis patients
must be maintained in a safe and functional working condition. The facility’s program for
Page 171 of 420
preventive maintenance and repair of all equipment must be in accordance with the equipment
manufacturer’s instructions.
Staff must operate and maintain the equipment in accordance with manufacturer’s instructions.
Malfunctioning machines awaiting repair must be removed from service and should be labeled
or tagged to prevent use. The facility should have a plan for the operation and routine
maintenance of at least the following equipment and equipment systems:
Hemodialysis delivery system:
• High flux dialyzers may only be used with machines specified by the manufacturer as
capable of accurately monitoring and controlling fluid removal.
• If heparin pumps are incorporated into the delivery system, the pumps must be
maintained as clean and functional.
• As required by manufacturers, testing of safety features, e.g., alarms, pressure
holding tests, and independent verification of dialysate pH and conductivity should be
conducted prior to each dialysis treatment.
“Dummy” drip chambers:
• “Dummy” drip chambers are fluid-filled chambers that are used to bypass the
dialysis machine’s air detectors.
• The practice of using a “dummy” drip chamber to prepare a dialysis machine for
patient use is hazardous to patient safety and risks the undetected infusion of air into
a patient if the “dummy” drip chamber is not removed at the initiation of dialysis.
• The presence and availability of “dummy” drip chambers in the hemodialysis patient
treatment area is considered an immediate and serious threat to patient health and
safety. Unintentional failure to remove the “dummy drip chamber” from the air
detector places the patient at risk for potential life-threatening events including air
embolism and other possible adverse effects as use of a dummy drip chamber during
a patient’s dialysis treatment would be inconsistent with the manufacturer’s DFU.
• Use of a dummy drip chamber on the treatment floor departs from the standards of
care for dialysis. The use of dummy drip chambers in dialysis facilities is referenced
in several clinical studies, and warnings against using dummy drip chambers in the
treatment area are included in nearly all manufacturers’ instructions. The use of
“dummy” drip chambers in the dialysis facility is typically reserved for machine
maintenance purposes and is only used outside of the patient treatment area.
Water treatment system:
Requirements for monitoring and maintenance of the water treatment system are incorporated
into the Condition for water and dialysate quality at § 494.40 and should be cited there.
Dialyzer reprocessing system:
Page 172 of 420
Requirements for monitoring and maintenance of dialyzer reprocessing systems are incorporated
into the Conditions for Reuse of Hemodialyzers and Bloodlines at § 494.50 and should be cited
there.
Ancillary equipment:
• Ancillary equipment may include, but is not limited to: functional and clean patient
scales, centrifuge, refrigerators, incubators for in-house performance of
water/dialysate cultures, emergency generators, blood pressure monitoring
equipment, infusion pumps, patient thermometers, eye wash stations, conductivity and
pH meters, Hoyer lifts, and equipment required to provide in-house laboratory testing
(e.g. blood glucose meters, heat blocks, equipment for activated clotting times (ACT),
supplies for testing for occult blood and hematocrit levels).
• Maintenance of refrigerators should include the monitoring of temperatures to ensure
these are appropriate for the items stored.
• If a generator is present, documentation should be available regarding testing and
maintenance per the manufacturer’s instructions.
• Records should be available regarding the daily cleaning, testing, and periodic
calibration of pH and conductivity meters as recommended by the manufacturer.
• Documentation of periodic calibration of patient scales, blood pressure devices,
blood volume monitors, and laboratory equipment, as applicable, should be
available.
Emergency equipment:
Emergency equipment as referenced at E-0028 of Appendix Z should be clean, functional, and
accessible. Use E-0028 for the lack of required emergency equipment; use this tag for failure to
maintain the emergency equipment.
Furniture:
• Patient treatment chairs, facility wheelchairs, and waiting area chairs must be
maintained to allow effective cleaning/disinfection.
• Torn upholstery must be repaired or replaced; broken mechanisms (e.g. footrests,
reclining levers) must be repaired or the equipment removed from use.
The facility equipment maintenance program should include documentation regarding all
equipment or devices used for home patients, whether maintained by the facility or by durable
medical equipment suppliers (DME). Refer to the Condition for Care at home at V597 if
problems with the maintenance and/or exchange of home dialysis equipment are identified.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
82efd5d57cc38074684eda1f26f6d06f79d392e6ba8d0ee0940abe99ef5753df
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.