US · guidance
CMS SOM App. H, Tag V339
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR §
494.50(b)(1)
11.4.1 Interior (blood/dialysate compartment)
11.4.1.1 Germicidal process
Chemical germicides or other procedures used for disinfecting of hemodialyzers have been
shown to accomplish at least high-level disinfection when tested in dialyzers artificially
contaminated with appropriate microorganisms.
If the germicide has an expiration date from the manufacturer, staff members should be
sure that the chemical is not outdated. Some germicides have recommendations for
maximum storage time after dilution or activation and before usage. If this is the case, the
expiration date of the prepared germicide solution should be marked on the outside of the
germicide solution container, and that date should be checked at the beginning of each day,
before reprocessing begins.
Page 146 of 420
The disinfection process shall not adversely affect the integrity of the dialyzer. Germicides
shall be rinsed from the dialyzer to below known toxic levels within a rinse-out period
established for the particular germicide (see AAMI 12.4). To prevent injury, staff members
shall take care not to mix reactive materials such as sodium hypochlorite and
formaldehyde.
Interpretive Guidance § 494.50(b)(1)
ANSI/AAMI RD47:2002/A1:2003
11.4.1 Interior (blood/dialysate compartment)
11.4.1.1 Germicidal process
If formaldehyde is used as the sole germicidal agent, the CDC recommends that a concentration
of 4% (W/V) be used in both the blood and dialysate compartments with a minimum contact time
of 24 hours at a temperature of at least 20° C; lower concentrations or shorter contact times are
appropriate if equivalent results can be demonstrated under other conditions. Formaldehyde
used for reprocessing dialyzers shall not be cloudy. Concentrated formaldehyde stored under
adverse conditions can polymerize to form paraformaldehyde, a white precipitate. Formaldehyde
should be of United States Pharmacopoeia (USP) standards or better quality. When other
germicides are used, the manufacturer’s instructions should be followed. If maximum storage
temperature limitations exist, records should be maintained to document this criterion.
AAMI Rationale for the Development and Provisions of this Recommended Practice
A.11.4.1.1 Germicidal process
Unfortunately, no realistic procedure exists whereby a dialysis center can monitor the
effectiveness of the disinfection procedure. Such sophisticated microbiological tests cannot be
performed in dialysis centers because they require the use of specialized equipment and highly
trained microbiologists. Due to these constraints, a center should adhere strictly to established
protocols for QC and QA. Tests for total bacteria and endotoxin in the water used to make up the
germicide should be conducted at least monthly. If there are problems in maintaining water
quality at the level established by ANSI/AAMI RD62:2001, Water treatment equipment for
hemodialysis applications (referenced in RD52:2004), the testing may need to be performed
more frequently.
Testing the germicide’s final-use concentration should be a part of the center’s QC program as
well as verifying that each dialyzer was filled with germicide.
Potency testing of each batch of germicide is specifically recommended for batches of manually-prepared germicides, regardless of whether they are used with a manual or an automated
system. Germicide solutions diluted online by automated machines should be checked for
concentration at least monthly.
CMS requires (at 42 CFR § 494.50) that dialyzers not be subjected to multiple germicide
solutions because of possible combined actions of the germicides on the hemodialyzer
membrane. This requirement does not apply to the original sterilization process or chemical
Page 147 of 420
cleaning agents to which the hemodialyzer may be exposed for short periods during the cleaning
process for reuse.
Additional Guidance:
The reuse staff must be knowledgeable about the germicide used and the risks this germicide
presents to him/her and to the patients, such as injuries that may be caused by mixing reactive
materials such as sodium hypochlorite and formaldehyde.
Containers of germicide should be dated to indicate dilution and discard dates.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
de017bb53073e7fd9a77bebb595e98ae8a0621b4d45116df10a784d0f8b40467
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.