CMS SOM App. H, Tag V325
ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR §
494.50(b)(1)
7.2.2 Process control testing: concentration of germicide
7.2.2.2
The test for the concentration of germicide or chemical shall be established before clinical
use of the reprocessed dialyzers ([AAMI] 11.4.1.6 and 12.3.2). For systems using heat
disinfection, verifiable evidence shall be available before the next use that dialyzers have
been exposed to the appropriate temperature for the time required. If chemicals are used
to enhance heat disinfection, both a presence test and a verification of time and
temperature shall be performed.
Interpretive Guidance § 494.50(b)(1)
The reuse manual should document how the concentration of germicide will be tested.
If heat disinfection is the reprocessing method, records of each batch of dialyzers processed must
include an indicator, such as an automated time/temperature recording log, that the dialyzers
were exposed to the appropriate temperature for the time required. If a chemical, such as citric
Page 134 of 420
acid, is used to enhance heat disinfection, a presence test is also required before clinical use of
the dialyzers.
If an incubator or oven is used to raise the dialyzer storage temperatures, a recording
thermometer should be in use to ensure sufficient temperature is consistently maintained.
Records should document that these devices functioned as expected.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
bb252a50b1bb367aea1ec95d7c7c258d1e347faa9f036b7b66aaefd4df0cfc71
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.