CMS SOM App. H, Tag V322

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR §

activein force · 2026-07-22 – presentas-observed

494.50(b)(1)

9 Reprocessing supplies

9.1 Specifications and testing

Each reprocessing material should meet a written specification. The fulfillment of that

requirement may be determined by certification by the product’s supplier that the product

meets necessary specifications, labeling for its intended purpose, or by testing procedures

by trained personnel, as appropriate. The requirement may also be complied with by

purchasing a specific grade as specified by the process, such as USP citric acid. When the

user performs testing, he or she should maintain a log of the date of test, the identifying

number (lot number) of the batch, the person performing any testing, and the test results.

When bleach is purchased from a commercial outlet, the labeled concentration should be

between 5.25% and 6.15%, and the formula should not contain fragrances or scents.

Interpretive Guidance § 494.50(b)(1)

ANSI/AAMI RD47:2002/A1:2003

9 Reprocessing supplies

9.1 Specifications and testing

Over the past few years, bleach (sodium hypochlorite) manufacturers have begun selling

household bleach in many new formulas. The concentration of sodium hypochlorite has gone

Page 132 of 420

from 5.25% to 6.15% in many cases. The CDC has not changed its recommendations for diluting

the bleach to take into account these percentage changes.

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.9 Reprocessing supplies

A.9.1 Specifications and testing

Testing of all incoming materials had been proposed. In recognition of the fact that most medical

supplies are certified by the vendor and not tested by the user, the AAMI RDD Committee

decided to recommend that supplies need not be tested by the facility doing hemodialyzer

reprocessing if they are marketed for hemodialyzer reprocessing.

Additional Guidance:

Most materials used in reprocessing will not require testing by the user since they are validated

and certified by a vendor. If the facility is using a material not commonly used in reprocessing, it

may not be certified by a supplier as meeting necessary specifications . For these materials,

there must be documentation of testing done by trained staff members to verify that the product

meets the necessary specifications.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
80ef05d9fc3c9385cfc82bd975fe6c25db80c039aa8ebada9971919ccd8a58c1
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.