CMS SOM App. H, Tag V314

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR §

activein force · 2026-07-22 – presentas-observed

494.50(b)(1)

7.1 Water systems

The system providing water for reprocessing shall meet all of the requirements for

pressure and flow rate for operating the reprocessing equipment under minimal and peak

load conditions. Product water used for rinsing, cleaning, filling, and diluting the germicide

shall be shown to comply with the chemical and microbiological quality requirements

[specified in these regulations].

Water bacteriology monitoring shall be carried out where the dialyzer is connected to the

reuse system or as close as possible to that point.

Page 126 of 420

11.4 Germicide

11.4.1.5 Water quality monitoring

The water used to rinse and clean dialyzers and dilute the germicide should be tested for

bacterial contamination and pyrogens according to the requirements [of these regulations]

before a reprocessing program is undertaken. Once dialysis with the reprocessed

hemodialyzers has begun, testing for bacterial contamination should be frequent (e.g.,

weekly). Less frequent testing, but not less than monthly, may be appropriate if there is a

documented history of at least 3 months of results consistently below the required levels.

Interpretive Guidance § 494.50(b)(1)

The product water chemical and microbiological requirements outlined in this section are the

same as those in ANSI/AAMI RD52:2004, incorporated by reference in these regulations under

the Condition for water and dialysate quality.

Water samples for microbial and endotoxin testing must be routinely taken each month from the

water supplying each reprocessing system, as close as possible to the point where the dialyzer

would be connected to the system. If more than one automated reprocessing system is in use, the

water supply to each system must be monitored monthly.

New facilities or facilities which add dialyzer reprocessing should validate the safety of the

water supply to the reprocessing system by testing for bacteria (microbial content) and pyrogens

(endotoxins) weekly for at least 3 months, and at least monthly thereafter. The sites to be tested

include the water supply to the sinks used for rinsing dialyzers, to outlets used for mixing

germicide, and to each reprocessing system.

Note: If water quality samples do not consistently meet AAMI standards, i.e., at least 3

consecutive months of testing results that meet required levels, the water quality testing is

expected to be performed more frequent than monthly.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
2eb6815e27f225251e47e41aa726a994700a26d27c77375328e132d95798ba60
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.