The system providing water for reprocessing shall meet all of the requirements for
pressure and flow rate for operating the reprocessing equipment under minimal and peak
load conditions. Product water used for rinsing, cleaning, filling, and diluting the germicide
shall be shown to comply with the chemical and microbiological quality requirements
[specified in these regulations].
Water bacteriology monitoring shall be carried out where the dialyzer is connected to the
reuse system or as close as possible to that point.
Page 126 of 420
11.4 Germicide
11.4.1.5 Water quality monitoring
The water used to rinse and clean dialyzers and dilute the germicide should be tested for
bacterial contamination and pyrogens according to the requirements [of these regulations]
before a reprocessing program is undertaken. Once dialysis with the reprocessed
hemodialyzers has begun, testing for bacterial contamination should be frequent (e.g.,
weekly). Less frequent testing, but not less than monthly, may be appropriate if there is a
documented history of at least 3 months of results consistently below the required levels.
Interpretive Guidance § 494.50(b)(1)
The product water chemical and microbiological requirements outlined in this section are the
same as those in ANSI/AAMI RD52:2004, incorporated by reference in these regulations under
the Condition for water and dialysate quality.
Water samples for microbial and endotoxin testing must be routinely taken each month from the
water supplying each reprocessing system, as close as possible to the point where the dialyzer
would be connected to the system. If more than one automated reprocessing system is in use, the
water supply to each system must be monitored monthly.
New facilities or facilities which add dialyzer reprocessing should validate the safety of the
water supply to the reprocessing system by testing for bacteria (microbial content) and pyrogens
(endotoxins) weekly for at least 3 months, and at least monthly thereafter. The sites to be tested
include the water supply to the sinks used for rinsing dialyzers, to outlets used for mixing
germicide, and to each reprocessing system.
Note: If water quality samples do not consistently meet AAMI standards, i.e., at least 3
consecutive months of testing results that meet required levels, the water quality testing is
expected to be performed more frequent than monthly.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
2eb6815e27f225251e47e41aa726a994700a26d27c77375328e132d95798ba60
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.