§ 494.50(b) - Reprocessing requirements for the reuse of hemodialyzers and bloodlines: A
dialysis facility that reuses hemodialyzers and bloodlines must adhere to the following
reprocessing guidelines:
(1) Meet the requirements of AAMI published in “Reuse of Hemodialyzers,” third
edition, ANSI/AAMI RD47:2002 and RD47:2002/A1:2003. The Director of the Federal
Register approves this incorporation by reference in accordance with 5 U.S.C. 552(a) and 1
CFR Part 51. This publication is available for inspection at the CMS Information Resource
Center, 7500 Security Boulevard, Central Building, Baltimore, MD or at the National
Archives and Records Administration (NARA). For information on the availability of this
material at NARA, call 202-741-6030, or go to:
http://code_of_regulations/ibr_locations.html. Copies may be purchased from the
Association for the Advancement of Medical Instrumentation, 3300 Washington
Boulevard, Suite 400, Arlington, VA 22201- 4598.
Interpretive Guidance § 494.50(b)
The AAMI “Reuse of Hemodialyzers,” third edition, ANSI/AAMI RD47:2002/A1:2003 is
incorporated by reference. The recommendations are provided in the “Regulation” column and
carry the full weight of regulation. Explanatory language from the AAMI recommendations and
portions of the AAMI rationale for the development and provisions of the recommended practice
are provided in the “Interpretive Guidance" column as an aid to understanding the regulation.
In some cases, the AAMI identification numbers may appear out of order, as they have been
rearranged for better flow with the survey process. While exact language from the AAMI
document has been incorporated as regulation, the language in the “Interpretive Guidance”
column has been edited for clarity, brevity, and to minimize redundancy.
Page 119 of 420
Note: When words are inserted or altered in a direct quotation, square brackets—[]—are
placed around the change. The brackets enclose words intended to explain the quote or to help
integrate the quote into the guidance.
AAMI RD47:2002/A1:2003
11 Reprocessing
The multiple use of a dialyzer begins with the labeling of the new dialyzer (see AAMI section 10)
and then continues with the reprocessing procedures described in this section. Preparation of the
reprocessed dialyzer for the next dialysis is described in AAMI section 12. The cycle is repeated
after the next use of the dialyzer until the dialyzer does not meet the criteria for continued use.
The results of the tests and the signature or other unique means of identifying the person
performing each step should be maintained in a permanent record (see AAMI section 4.2).
Completion of all reprocessing steps, tests, and inspections should be documented in the
reprocessing record, accompanied by the signature or other unique means of identification of the
person completing them. When appropriate for the reprocessing procedure in use, all dialyzer
manufacturer’s instructions regarding reuse should be carefully followed.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
ef1401d59a4464eb240d33b4df2d861c1f299abd319cc571e42b0f56619a1507
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.