US · guidance
CMS SOM App. H, Tag V274
§ 494.40(c) - Corrective action plan
Water testing results including, but not limited to,
chemical, microbial, and endotoxin levels which meet AAMI action levels or deviate from
the AAMI standards must be addressed with a corrective action plan that ensures patient
safety.
Interpretive Guidance § 494.40(c)
Facility policy must include requirements to develop and implement a corrective action plan that
ensures patient safety when water testing results exceed acceptable limits or action levels.
A corrective action plan must be developed and implemented to ensure patient safety if any of the
results of water chemical analysis or microbial and endotoxin testing for water and dialysate
exceed the AAMI action levels or fall outside the AAMI standards. Water and dialysate
monitoring must be reported in the QAPI materials, and the medical director must be involved in
analyzing and addressing test results outside of expected parameters.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
2c24f5894e920b6263727273c98b36ddc15d6c950530b0c7319b700aef9d57af
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.