CMS SOM App. H, Tag V252

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

activein force · 2026-07-22 – presentas-observed

7.2 Microbial monitoring methods

7.2.1 General

7.2.2 Sample collection

7.2.1 General

Culture water …weekly for new systems until a pattern has been established. For

established systems, culture monthly unless a greater frequency is dictated by historical

data at a given institution.

Monitoring can be accomplished by direct plate counts, in conjunction with the

measurement of bacterial endotoxin.

7.2.2 Sample collection

Water samples should be collected directly from outlet taps situated in different parts of

the water distribution system. In general, the sample taps should be opened and the water

should be allowed to run for at least 60 seconds before a sample is collected in a sterile,

endotoxin-free container. A minimum of 50 mL of water, or the volume specified by the

laboratory performing the test, should be collected. Sample taps should not be disinfected.

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

7.2.2 Sample collection

Page 106 of 420

All new sterile plastic ware is endotoxin-free because of the high temperatures involved in the

manufacturing process.

Additional Guidance:

Bacterial colony counts may be done by an outside lab (thus the reference to direct plate counts)

or by the use of dip samplers, as described at V256. If a pattern of results greater than the

accepted limits is identified, the frequency of cultures (and subsequent disinfection) should be

increased to reach a frequency that ensures the system is being maintained below the allowed

contamination level between scheduled sampling. Refer to V255.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
c6b0e85345315d458da42415226a40736d8fb5004b5749b76cc303d66ea1aa7a
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.