US · guidance
CMS SOM App. H, Tag V250
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.6 Dialysate proportioning (dialysate conductivity and pH measurement)
[I]t is necessary for the operator to follow the manufacturer’s instructions regarding
dialysate conductivity and to measure approximate pH with an independent method before
starting the treatment of the next patient.
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.6 Dialysate proportioning
Injuries related to improper dialysate are rare, but they can and do happen when all procedures
are not followed. Frequently, when the error occurs, several patients have been exposed before
the facility recognizes the mistake. For example, because one of the concentrates is quite acidic
and the other is basic, connecting the wrong concentrates to the machine could result in
dialysate that could harm the patient.
Even though a single concentrate is used to prepare acetate dialysate, conductivity and pH
should be checked, because certain mix-ups involving acid concentrate and other chemicals can
result in an acceptable conductivity and very low pH.
Page 103 of 420
Additional Guidance:
Each machine must be tested for pH using a handheld meter or other appropriate testing device
(i.e., adequately sensitive testing strips) before every dialysis treatment and whenever a different
composition of acid concentrate is used. If the dialysis machine manufacturer requires testing of
conductivity, this must also be tested using an independent testing device prior to each treatment
and before using a different composition of acid concentrate during the same treatment. The
facility must have set limits for the allowable variability of the handheld values from the machine
readings. As new technology evolves, the manufacturer’s guidance related to this requirement
should be followed.
No Tag
(Rev. )
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
6 Monitoring
6.1 General
Quality control and quality assurance procedures should be established to ensure ongoing
conformance to policies and procedures regarding dialysate quality. This clause defines
some of the monitoring activities to be conducted at the dialysis facility as part of the
quality assurance process. The test methods described in [AAMI] 6.2 do not represent the
only acceptable methods available, but are intended to provide examples of acceptable
methods. The frequency of monitoring is generally recommended by the equipment
manufacturer. Table 4 can be used as a guideline for setting up a quality assurance
monitoring program in the absence of a manufacturer’s recommendations or to
supplement those recommendations.
Interpretive Guidance § 494.40(a)
This is an informational tag. Expected monitoring is listed under each water system and
dialysate component; there may be some variation from this Table based on specific equipment
in use.
ANSI/AAMI RD52:2004
6 Monitoring
6.1 General
Table 4—Monitoring guidelines for water purification equipment and distribution systems
and dialysate
NOTE—Refer to note below the table for an explanation of the use of Xs in the Specification
column.
Page 104 of 420
Item to monitor What to monitor Special
interval
Normal interval Specification
Sediment filter Pressure drop across the
filter
NA Daily Pressure drop less than XXXX
Sediment filter
backwashing
cycle
Backwash cycle timer
setting
NA Daily—beginning
of the day
Backwash clock set to XX:XX
Cartridge filter Pressure drop across
the filter
NA Daily Pressure drop less than XXXX
Water softener Product water
softness
NA Daily—end of the
day
Hardness as calcium
carbonate
less than 1 grain/gal, unless
otherwise specified by the
manufacturer of the reverse
osmosis equipment
Water softener
brine tank
Level of undissolved
salt in tank
NA Daily—end of the
day
Salt level at XXX
Water softener
regeneration
cycle
Regeneration cycle
timer setting
NA Daily—beginning
of the day
Softener timer set to XX:XX
Carbon
adsorption
beds
Product water free
chlorine and/or total
chlorine between the
beds
NA Prior to beginning
each patient shift
< 0.1 mg/L of total chlorine
Chemical
injection
system
Level of chemical in
the reservoir, injector
function, value of the
controlling parameter
(e.g., pH)
NA Daily Chemical level in reservoir
≥ XXX; controlling parameter
in
range XX–XX
Reverse
osmosis
Product water
conductivity, total
dissolved solids
(TDS), or resistivity
and calculated
rejection
NA According to the
manufacturer’s
recommendations
(continuous
monitors)
Rejection ≥ XX%
Reverse
osmosis
Product and reject
flow rates, and
calculated recovery
NA Daily (continuous
monitors)
Product water flow rate
> X.X gpm; recovery in the
range XX–XX %
Deionizers Product water
resistivity
NA Continuous Resistivity > 1 megohm-cm
Ultrafilters Pressure drop across
the filter
NA Daily Pressure drop less than XXXX
Water storage
Tanks
Bacterial growth and
pyrogens
Weekly, until a
pattern of
consistent
compliance with
limits can be
demonstrated
NA Bacterial count ≤ 50
CFU/mL;
endotoxin ≤ 1 EU/mL
Water
distribution
piping system
Bacterial growth and
pyrogens
Weekly, until a
pattern of
consistent
compliance with
limits can be
demonstrated
Monthly Bacterial count ≤ 50
CFU/mL;
endotoxin ≤ 1 EU/mL
UV light
sources
Energy output NA Monthly Light output > XXX
Ozone
generators
Concentration in the
water
NA During each
disinfection
Ozone concentration > XXX
Hot water
disinfection
systems
Temperature and
time of exposure of
the system to hot
NA During each
disinfection
Temperature not less than
Page 105 of 420
Item to monitor What to monitor Special
interval
Normal interval Specification
water XX °C; minimum exposure
time at temperature ≥ XX
minutes
Dialysate Bacterial growth and
endotoxin in the
dialysate
NA Monthly, rotated
among machines
so that at least
two machines are
tested each month
and so that each
machine is tested
at least once per
year
Bacterial growth ≤50
CFU/mL;
endotoxin ≤ 1 EU/mL
Dialysate Conductivity and pH NA Each treatment Conductivity within ± 5% of
the nominal machine value;
pH in the range 6.9–7.6
NOTE: It is not possible to specify universally acceptable operating ranges for each device listed
in the table, since some of these values will be system-specific. In those cases (denoted by Xs in
the Specification column of the table), the facility should define an acceptable operating range
based on the manufacturer’s instructions or measurements of system performance.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
ec5d28c210e7273aba09ab8286743c90cd51cc626e6997571fb9cd64a8b44a11
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.