ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
6.5 Concentrate distribution (Initial Bicarbonate Monitoring)
Page 97 of 420
Once a bicarbonate distribution system has been activated, dialysate should be monitored
weekly until sufficient data has been obtained to demonstrate consistent compliance with
acceptable levels of contamination. The frequency of monitoring may then be reduced, but
monitoring should be performed at least monthly. If elevated bacteria or endotoxin levels
are found in the dialysate, all systems involved in dialysate preparation, including the
bicarbonate concentrate distribution system should be evaluated and appropriate action,
such as disinfection, should be taken. The frequency of monitoring should then be
increased until it can be demonstrated that the problem has been resolved.
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
6.5 Concentrate distribution
Because acid concentrate distribution systems have been shown not to be subject to bacterial
proliferation, it is not necessary to perform bacteria and endotoxin testing on those systems.
Additional Guidance:
Whenever use of a new bicarbonate distribution system is initiated, weekly monitoring of
dialysate should occur for at least four consecutive weekly reports of acceptable levels.
Evaluation of positive culture or endotoxin reports should also consider the number of positives
in relationship to the number of samples taken. For example, if one sample out of 10 has a count
of 53, while the other 9 have no growth, repeating the culture of one or more sites may be the
first action taken. The medical director should be involved in any decisions for determining the
appropriate action(s) to address abnormal culture or endotoxin results, i.e., growth exceeds
permissible standards.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
a377a05ec6f9a7bb809d1576a84d9af0cb44d520fe2a1e7b008f0fca76f62303
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.