CMS SOM App. H, Tag V223
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.4 Concentrate preparation
5.4.2 Materials compatibility
All components used in concentrate preparation systems (including mixing and storage
tanks, pumps, valves, and piping) shall be fabricated from materials (e.g., plastics or
appropriate stainless steel) that do not interact chemically or physically with the
concentrate so as to affect its purity, or with the germicides or germicidal procedure used
to disinfect the equipment. The use of materials that are known to cause toxicity in
hemodialysis, such as copper, brass, galvanized material, and aluminum, are specifically
prohibited.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
cc876dc30a11657039ed1132564df6916112279448af1a08fd5402547e9a94c6
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.