ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

activein force · 2026-07-22 – presentas-observed

5.3.4.2 Ozone generators: system requirements/monitoring

6.3.4 Bacterial control devices

Ozone can be used for bacterial control only in systems constructed from ozone-resistant

materials (see AAMI 5.3.3 for suitable piping materials).

5.3.4.2 Ozone generators

Refer to RD62:2001, 4.3.15 Ozone disinfection systems: When used to control bacterial

proliferation in water storage and distribution systems, an ozone generator should be

capable of delivering ozone at the concentration and for the exposure time specified by the

manufacturer.

6.3.4 Bacterial control devices

6.3.4.2 Ozone generators

Ozone generators should be monitored for ozone output at a level specified by the

manufacturer. The output of the ozone generator should be measured by the ozone

concentration in the water. A test based on indigo trisulfonate chemistry, or the equivalent,

should be used to measure the ozone concentration …each time disinfection is performed.

An ozone-in-ambient-air test should be conducted on a periodic basis, as recommended by

the manufacturer, to ensure compliance with the OSHA permissible exposure limit of 0.1

ppm. A log sheet should be used to indicate that monitoring has been performed.

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.3.4.2 Ozone generators

Ozone may be used to control bacterial proliferation in water storage and distribution systems.

Ozone may also degrade endotoxins. Therefore, ozone can be used for bacterial control only in

systems constructed from ozone-resistant materials. Ozone generators convert oxygen in air to

ozone using a corona discharge or ultraviolet irradiation. The ozonated air is then injected into

Page 80 of 420

the water stream. An ozone concentration of 0.2 mg/L to 0.5 mg/L, combined with a contact time

of 10 minutes, is capable of killing bacteria, bacterial spores, and viruses in water. Destruction

of established biofilm may require longer exposure times and/or higher concentrations of ozone.

Ozone may degrade many plastic materials, including PVC and elastomeric O-rings and seals.

Additional Guidance:

Ozone exposure may have a harmful and destructive effect on some plastic or metal materials.

Therefore, it is not recommended for disinfection of many of the water distribution loop

materials. Refer to Table 2 at V212.

The manufacturer of an ozone disinfection system is typically responsible for validating the

recommended operating conditions to demonstrate that the system provides adequate reduction

in bacterial and endotoxin levels. Refer to the manufacturer’s guidance for required

concentrations, contact time, and disclosure on the effectiveness of the system in killing bacteria

and endotoxins.

The presence of ozone in product water may be harmful to patients. Staff must monitor the

system disinfection to ensure that (1) the generator delivers ozone at the concentration and

exposure time specified by the manufacturer and (2) there is a mechanism in place to detect

ozone in the area of the ozone generator when ozone disinfection systems are used. Facility

policy must define the frequency for “periodic basis” for testing the levels of ozone in ambient

air. This policy should reflect the manufacturer’s guidance and OSHA requirements for exposure

limits.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
fb4a9621b3abbc3beead68d16b8a060abdc42f2012c446b3948cba5c8d482bd4
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.