CMS SOM App. H, Tag V216
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.3.4.2 Ozone generators: system requirements/monitoring
6.3.4 Bacterial control devices
Ozone can be used for bacterial control only in systems constructed from ozone-resistant
materials (see AAMI 5.3.3 for suitable piping materials).
5.3.4.2 Ozone generators
Refer to RD62:2001, 4.3.15 Ozone disinfection systems: When used to control bacterial
proliferation in water storage and distribution systems, an ozone generator should be
capable of delivering ozone at the concentration and for the exposure time specified by the
manufacturer.
6.3.4 Bacterial control devices
6.3.4.2 Ozone generators
Ozone generators should be monitored for ozone output at a level specified by the
manufacturer. The output of the ozone generator should be measured by the ozone
concentration in the water. A test based on indigo trisulfonate chemistry, or the equivalent,
should be used to measure the ozone concentration …each time disinfection is performed.
An ozone-in-ambient-air test should be conducted on a periodic basis, as recommended by
the manufacturer, to ensure compliance with the OSHA permissible exposure limit of 0.1
ppm. A log sheet should be used to indicate that monitoring has been performed.
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.3.4.2 Ozone generators
Ozone may be used to control bacterial proliferation in water storage and distribution systems.
Ozone may also degrade endotoxins. Therefore, ozone can be used for bacterial control only in
systems constructed from ozone-resistant materials. Ozone generators convert oxygen in air to
ozone using a corona discharge or ultraviolet irradiation. The ozonated air is then injected into
Page 80 of 420
the water stream. An ozone concentration of 0.2 mg/L to 0.5 mg/L, combined with a contact time
of 10 minutes, is capable of killing bacteria, bacterial spores, and viruses in water. Destruction
of established biofilm may require longer exposure times and/or higher concentrations of ozone.
Ozone may degrade many plastic materials, including PVC and elastomeric O-rings and seals.
Additional Guidance:
Ozone exposure may have a harmful and destructive effect on some plastic or metal materials.
Therefore, it is not recommended for disinfection of many of the water distribution loop
materials. Refer to Table 2 at V212.
The manufacturer of an ozone disinfection system is typically responsible for validating the
recommended operating conditions to demonstrate that the system provides adequate reduction
in bacterial and endotoxin levels. Refer to the manufacturer’s guidance for required
concentrations, contact time, and disclosure on the effectiveness of the system in killing bacteria
and endotoxins.
The presence of ozone in product water may be harmful to patients. Staff must monitor the
system disinfection to ensure that (1) the generator delivers ozone at the concentration and
exposure time specified by the manufacturer and (2) there is a mechanism in place to detect
ozone in the area of the ozone generator when ozone disinfection systems are used. Facility
policy must define the frequency for “periodic basis” for testing the levels of ozone in ambient
air. This policy should reflect the manufacturer’s guidance and OSHA requirements for exposure
limits.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
fb4a9621b3abbc3beead68d16b8a060abdc42f2012c446b3948cba5c8d482bd4
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.