ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

activein force · 2026-07-22 – presentas-observed

5.3.3 Water distribution systems (continuous flow rates)

7 Strategies for bacterial control

7.1 General

5.3.3 Water distribution systems

Water distribution systems should be configured as a continuous loop and designed to

minimize bacterial proliferation and biofilm formation. A centrifugal pump made of inert

materials is necessary to distribute the purified water and aid in effective disinfection.

7 Strategies for bacterial control

7.1 General

To minimize biofilm formation, there should always be flow in a piping system. A

minimum velocity of 3 ft/sec in the distal portion of the loop of an indirect feed system and

a minimum velocity of 1.5 ft/s in the distal portion of a direct feed system are recommended

when the system is operating under conditions of peak demand.

Dead-end pipes and unused branches and taps that can trap fluid must be eliminated

because they act as reservoirs of bacteria and are capable of continuously inoculating the

entire volume of the system. These measures also minimize the possibility that pockets of

residual disinfectant could remain in the piping system after disinfection.

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.3.3 Water distribution systems

A multistage centrifugal pump is preferred for this purpose.

7 Strategies for bacterial control

7.1 General

Other measures can also help protect pipes from contamination.

Additional Guidance:

It is not the intent of this regulation to dictate the type of pump used for this purpose, other than

that the pump be made of inert material.

Page 74 of 420

It is not intended that surveyors would measure flow rates; responsible staff (i.e., the chief

technician) should be able to describe how the system is monitored to ensure at least minimum

flow, i.e. constant flow passing through the piping system, during operation.

The facility must identify and remove any potential dead-end pipes. For example, if the facility

has discontinued reprocessing dialyzers, any unused piping should be removed to prevent

stagnant water or trapping of disinfectant.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
1cd049dc54fc88fcf6fec64c2a6d6d9d57c9a0e85d6dc5b1b932392a4026e7da
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.