ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

activein force · 2026-07-22 – presentas-observed

5.2.8 Deionization (alarms/divert to drain)

An audible and visual alarm shall be activated when the product water resistivity falls

below this level and the product water stream shall be prevented from reaching any point

of use, for example by being diverted to drain. The alarm must be audible in the patient

care area. [ANSI/AAMI RD52:2004 guidance directs users to RD62:2001, Section 4.3.6]

ANSI/AAMI RD52:2004

5.2.8 Deionization

Specifically, the resistivity monitor following the final deionizer bed shall be connected to

an audible and visible alarm in the dialysis treatment area, and the DI system shall divert

product water to drain or otherwise prevent product water from entering the distribution

system should an alarm condition occur. Under no circumstances shall DI be used when the

product water of the final bed has a resistivity below 1 megohm-cm.

Interpretive Guidance § 494.40(a)

Page 67 of 420

ANSI/AAMI RD52:2004

5.2.8 Deionization

DI has a finite capacity that, when exceeded, will cause dangerously high levels of contaminants

in the product water. Fortunately, the quality of product water from DI is easily monitored by

resistivity monitors that, when used as specified (e.g., minimum resistivity 1 megohm-cm or

greater), can prevent inadvertent operation of an exhausted deionization system.

Additional Guidance:

Except for home patients, there must be capability for either an automatic divert-to-drain system

or other means to prevent product water from entering the distribution system for any DI system

in use. DI has a very limited capacity that, when exceeded, will cause dangerously high levels of

contaminants in the product water. Due to the increased risk to patient health and safety with the

use of water from an exhausted DI tank, manually stopping the flow before the RO product water

reaches the patient may not be as sufficient as the use of an automated protection system, such as

a divert-to-drain valve.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
f06a195599e2d386c05deac1895c9dd12ffac34ea8bc39dfd517f9a67ec9d753
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.
CMS SOM App. H, Tag V203 — ANSI/AAMI RD52:2004 Requir… · binding.law