US · guidance
CMS SOM App. H, Tag V202
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.2.8 Deionization: (continuous monitor resistivity)
6.2.8 Deionization (monitoring)
Deionization: Deionization systems, when used to prepare water for hemodialysis
applications, shall be monitored continuously to produce water of one megohm/cm or
greater specific resistivity (or conductivity of one microsiemen/cm or less) at 25°C.
[Guidance for deionization systems monitoring in ANSI/AAMI RD52:2004 directs users to
ANSI/AAMI RD62:2001; Section 4.3.6]
5.2.8 Deionization
Page 65 of 420
Deionization may be used to polish product water from a reverse osmosis system or may be
used as a standby if the reverse osmosis system fails.
6.2.8 Deionization
Deionizers shall be monitored continuously using resistivity monitors that compensate for
temperature and are equipped with audible and visual alarms. Resistivity monitors shall
have a minimum sensitivity of 1.0 megohm-cm. Patients shall not be dialyzed on deionized
water with resistivity less than 1.0 megohm-cm measured at the output of the deionizer
Resistivity monitor readings should be recorded on a log sheet twice each treatment day.
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.2.8 Deionization
Deionization (DI) is an ion exchange process that removes both anions (negatively charged ions)
and cations (positively charged ions) from water. During the exchange process, hydroxyl ions
replace other feed water anions, and hydrogen ions replace other feed water cations; the
hydroxyl and hydrogen ions then combine to form pure water.
Water treated by DI may be very high quality with regard to the absence of ionized
contaminants, but the process does not remove nonionized substances, including bacteria and
bacterial endotoxins. DI systems may contain anion and cation resin in separate vessels, known
as “dual-bed systems,” or may have both resin types mixed together in a single vessel, known as
“mixed-bed” or “unibed systems.”
AAMI Rationale for the Development and Provision of this Recommended Practice
A.5.2.8 Deionization
Deionizers are an effective means of removing ionic contaminants from water. However, they do
not remove nonionic species (such as bacteria), and they may contribute bacterial contaminants
to the water rather than remove them. The inability of deionizers to remove nonionic
contaminants may limit the removal of aluminum by deionization. Deionizers have a finite
capacity for contaminant removal. Once the deionizer is depleted of hydrogen and hydroxyl ions,
the next least avidly bound ions will be displaced by more avidly bound ions. For example, once
the hydroxyl ions are depleted, anionic contaminants in the water will displace fluoride ions
from the anion exchange resin.
This phenomenon has led to high levels of fluoride in the product water, with subsequent patient
injury and/or death. For the above reasons, the use of deionization as the primary means of
purification is strongly discouraged. Deionization may be used to polish product water from a
reverse osmosis system or may be used as a standby if the reverse osmosis system fails.
Deionizers offer a large surface area for bacterial proliferation and deionizers generally
contribute to the bioburden in the water. The tendency for deionizers to contribute bacterial
contaminants to the water is greater when deionizers are kept as a backup for a reverse osmosis
system, particularly if there is no flow through the deionizers. Some facilities counter this
Page 66 of 420
tendency by connecting the deionizers in parallel to the main water line and by maintaining a
low flow through them. An alternative approach is to contract with a local vendor to provide
backup deionizers on demand.
Suggested deionizer monitoring guidelines from ANSI/AAMI RD52, Table 4 include:
Product water resistivity must be continuously monitored, with a result of resistivity >1
megohm-cm.
Additional Guidance:
If DI tanks are available for back-up use, the facility should take action to counter the tendency
of DI to contribute bacterial contaminants to the water. This may be accomplished by either
storing the tanks dry, placing the tanks on line post-RO so that there is a low flow of water
through them, or flushing the DI tanks daily. DI tanks should not be stored “wet,” i.e., filled with
stagnant water.
Exhausted DI tanks (<1.0 megohm-cm) present a serious risk to patients, and use of exhausted
DI tanks have resulted in deaths. If the water system uses DI as primary purification or as a
polish, the system must be closely monitored by knowledgeable staff. Pure water has a resistivity
of 18.3 megohms. Documentation of a reading greater than 18.3 megohms would indicate some
error. Exhausted DI tanks should be returned to the vendor for recharging. The date of exchange
should be posted on the tank(s) and recorded in a log.
Deionization is not required in every facility; the source water should determine the water
treatment components needed. If deionization is in use, the facility must follow these
requirements.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
69e33fdb26ecf3568a1cccb52957e2fdc2dcf549113ac53210c4efc4afb43371
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.