ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

activein force · 2026-07-22 – presentas-observed

6.2.7 Reverse osmosis: (Chemical analysis: frequency)

Chemical analysis for the contaminants listed in 4.1.1 (Table 1) should be done when the

RO system is installed, when membranes are replaced, and at not less than annual intervals

thereafter to ensure that the limits specified in 4.1.1 are met (see Table 1). Chemical

analyses should be done when seasonal variations in source water suggest worsening

quality or when rejection rates fall below 90%.

Interpretive Guidance § 494.40(a)

If your State has more stringent requirements, those must be followed to avoid potential

noncompliance with 42 C.F.R. § 494.20..Accordingly, facilities who fail to comply with state and

local laws would be cited at V100 or V101.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
44bc5e57eda19d7526e62ebda65a99c94eb70477b55972c5d441651053fe78b7
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.