ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40 (a)
6.2.4 Softeners: (Testing hardness/log)
Users should ensure that test accuracy and sensitivity are sufficient to satisfy the total
hardness monitoring requirements of the reverse osmosis machine manufacturer. Total
hardness of the water exiting the water softener should be measured at the end of each
treatment day.
Water hardness test results should be recorded in a water softener log.
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
6.2.4 Softeners
Page 51 of 420
Softener monitoring, which should be performed daily, involves testing effluent water for total
hardness to ensure that the limits established by the reverse osmosis machine manufacturer are
not exceeded. In the case of automatically regenerating softeners, monitoring also includes
verification that the brine tank contains a sufficient supply of undissolved sodium chloride and
that the control valve timer, when present, indicates the correct time of day.
Testing for hardness should be performed using an ethylenediaminetetracetic acid (EDTA)
titration test, with “dip and read” test strips, or a similar method.
The hardness test conducted at the end of the day will indicate the overall effectiveness of the
water softener under worst-case conditions and ensure that the softener is sized correctly—that
is, it has sufficient capacity, expressed in grains of calcium carbonate.
Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include:
For softeners: monitor product water softness daily at the end of the treatment day for hardness
as calcium carbonate, <l grain/gal, unless otherwise specified by the manufacturer of the
reverse osmosis equipment.
Additional Guidance:
The water softener log may be incorporated as a part of another log kept for the water treatment
system. For hardness tests requiring color differentiation, the person performing the analysis
should be capable of correctly interpreting the results, including the ability to distinguish and
accurately identify any colorimetric or other readings. If the person cannot differentiate these
colors, an automated meter should be used.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
38365a0cf83c6ea88b7fd242da5c706b48df8afb95f907d71c2cdd807fa183dc
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.