ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40 (a)

activein force · 2026-07-22 – presentas-observed

5.2.3 Cartridge filters (equipment configuration)

6.2.3 Cartridge filters (monitoring)

5.2.3 Cartridge filters

The cartridge is contained within an opaque filter housing with seals to separate the feed

and product water streams.

When the maximum [pressure drop] ∆P recommended by the filter manufacturer is

reached, the cartridge should be replaced according to the manufacturer’s instructions.

6.2.3 Cartridge filters

Cartridge filters should be monitored on a periodic…. basis for a [pressure drop] ∆P across

the filter [that] can be used to determine when the filter is retaining particulate matter to

the point that the filter will no longer allow the required water flow without an excessive

reduction in pressure at the outlet of the filter. A marked decrease in ∆P without a

corresponding decrease in flow rate may indicate a loss of filter integrity. Follow the

manufacturer’s recommendations concerning when to replace cartridge filters.

Replacement of the cartridge will usually be indicated by an increase in ∆P to some

specified value. A log sheet should be developed to record the pressure drop measurements.

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.2.3 Cartridge filters

Cartridge filters consist of a cylindrical cartridge of the filter medium with a central drainage

core. Although cartridge filters may be installed at the inlet to a water system, their usual

application is as a final filtration step before reverse osmosis.

As the cartridge accumulates particulate material, resistance to flow through the filter increases,

as indicated by an increase in ∆P.

Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include:

For cartridge filters: monitor for pressure drop across the filter daily, looking for a pressure

drop less than a number determined by facility policy. Since some values are system-specific, the

facility should define an acceptable operating range for monitoring pressure drops based on the

manufacturer’s instructions or measurements of system performance.

Additional Guidance:

Page 49 of 420

Cartridge filters are not required in every facility; the source water should determine the

necessary water treatment components. If cartridge filters are in use, the facility must follow

these requirements.

A marked decrease in ΔP could indicate that there is no filter in the container.

History

Rev.

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
e768778e16bbf60a7dc898ee73108a7df2ce86dfb8f30cfa140174334a026cda
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.