CMS SOM App. H, Tag V182
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5 Equipment
5.1 General:
A dialysis facility should develop contingency plans to cover the failure of its water
purification and distribution system or a critical component of that system. Such
contingency plans should describe how to deal with events that completely prevent dialysis
from being performed, such as failure of the facility’s municipal water supply or electrical
service following a natural disaster or water main break. Other plans should address how
to deal with sudden changes in municipal water quality, as well as with the failure of a
critical component of the water purification and distribution system.
Interpretive Guidance § 494.40(a)
Page 42 of 420
ANSI/AAMI RD52:2004
5 Equipment
5.1 General
This section on equipment provides a brief description of the different components that may be
included in a water purification and distribution system used for hemodialysis applications.
Since feed water quality and product water requirements may vary from facility to facility, not
all of the components described will be necessary in every purification and distribution system.
Routine dialysis requires a well-functioning water purification and distribution system, since
dialysis cannot be performed without an adequate supply of water. In addition, specific
components of the water purification and distribution system are critical to its operation. An
example of such a critical component is the circulating pump in an indirect feed system.
Additional Guidance:
An emergency or contingency “plan” that specifies the facility will use tap water or dechlorinated tap water may include evidence that the water intended for use has been found safe
for such use (i.e., has levels below AAMI accepted limits of aluminum, copper, chloramines,
fluoride, nitrate, sulfate, zinc and other contaminates known to be toxic to dialysis patients).
Refer also to the requirements for emergency policies and procedures found under the Condition
for Physical Environment at V408.
No Tag
(Rev. )
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.2 Water purification systems
5.2.1 General
Water purification systems consist of three basic sections: a pretreatment section that
conditions the water supplied to the primary purification device, which may be followed by
other devices that polish the final water quality. The pre-treatment section commonly
includes a sediment filter, cartridge filters capable of retaining particles of various sizes, a
softener, and carbon adsorption beds. The primary purification process most commonly
used is reverse osmosis, which may be followed by deionization and ultrafiltration for
polishing the product water from the reverse osmosis system.
Whether a particular device is included in an individual water purification system will be
dictated by local conditions.
Interpretive Guidance § 494.40(a)
AAMI Rationale for the Development and Provision of This Recommended Practice
A.5 Equipment
A.5.2 Water purification systems
Devices marketed to purify water for hemodialysis are also subject to the U.S. Food and Drug
Administration’s (FDA) 510(k) approval process. The FDA has published guidelines for water
Page 43 of 420
purification devices used in hemodialysis (Guidance for the Content of Premarket Notifications
for Water Purification Components and Systems for Hemodialysis). Water purification devices
marketed for use in hemodialysis applications must be approved by the FDA, and users should
ensure that devices obtained from vendors have been approved by the FDA.
Design and instrumentation of individual purification devices may vary from these general
descriptions. For example, softeners may be configured as a single resin bed that is regenerated
outside the regular operating hours of the dialysis unit, or they may have a dual-bed
configuration that allows one bed to be regenerated. At the same time, the other is used to
provide water for normal dialysis operations.
Depending on the feed water quality and product water requirements, not every component may
be required in a given facility. Likewise, additional components may be required in certain
circumstances. For example, carbon adsorption may not provide adequate chloramine removal
if the water contains substances, such as polyphosphates, that mask the reactive sites on the
carbon particles. In those circumstances, other processes, such as infusion of sodium
metabisulfite, may be required to achieve product water that meets these requirements.
Users are encouraged to obtain detailed descriptions of all purification processes, together with
operating manuals and maintenance procedures, from the manufacturer or the vendor providing
the water purification and distribution system.
Additional Guidance:
This is an informational tag; specific requirements for various water treatment components and
deficient practices should be cited under the applicable component tags.
Under FDA regulations at the time these regulations were published (April 15, 73FR20369), all
water treatment devices and systems installed after May 30, 1997 must meet review requirements
under section 510(k) of the Food, Drug, and Cosmetic Act (21 U.S.C. sec. 360(k)) as described
in Guidance for the Content of Premarket Notifications for Water Purification Components and
Systems for Hemodialysis. If a facility is found to be using water treatment devices installed after
1997 that have not been approved by the FDA, they may be found out of compliance with 42
C.F.R. §494.20. Equipment installed prior to 1997 is not required to have evidence of FDA
510(k) approval. Regardless of when a water purification system was installed, the system must
yield water and dialysate that meets these AAMI standards and must be monitored and
maintained in accordance with the ANSI/AAMI RD52 guidelines, as incorporated by reference in
these regulations.
Each piece of water treatment equipment described below is not required for every system, and
determinations of necessary equipment will be made by the facility based on a number of factors
such as feed water analysis and geographical considerations. The water treatment system must
be designed to process the local source water into AAMI quality water. All facilities should have
primary and secondary carbon tanks to remove chlorine/chloramines, and reverse osmosis (RO)
or deionization (DI) treatment systems. Any component in use should meet the requirements
specified in these rules for that component.
Page 44 of 420
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
67e4916fb93606fa40e9cf30cab00b9c6c2890c2a6d114e647cb7d8e53e85387
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.