ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
4.1.2 Bacteriology of water
The manufacturer or supplier of a complete water treatment and distribution system
[must] demonstrate that the complete water treatment, storage, and distribution system is
capable of meeting these requirements at the time of installation.
Following the installation of a water treatment, storage, and distribution system, the user is
responsible for continued monitoring of the water bacteriology within the system and for
complying with the requirements of this standard, including those related to action levels.
Page 39 of 420
Interpretive Guidance § 494.40(a)
While this standard is primarily addressed to the manufacturer of equipment, each dialysis
facility’s medical director is responsible for the quality of medical care provided to patients. The
medical director oversees the installation and monitoring of the water treatment and distribution
system. Their failure to evaluate and validate the environment where devices and equipment will
be used may result in non-compliance with this requirement.
Existing facilities must monitor AAMI water chemical analysis and microbial testing and take
action if the results fall outside AAMI standards. If culture results exceed the action levels, the
facility may need to repeat cultures or disinfect the system and repeat cultures (depending on the
number of positive cultures, etc.) and continue treatment while awaiting the results.
AAMI recommendations at 4.1.1, Table 1, recommend maximum levels of contaminants above
which corrective measures should be taken. Testing should occur initially upon installation, and
at annual intervals thereafter. If the water supply for the facility is from a private well, an annual
analysis of the product water quality may not be sufficient to ensure that the feed water
requirements of the water treatment system in use are continuously met. The quality of water
from the well may change over time, and private wells are not routinely monitored. More
frequent analysis may be needed if the well is subject to seasonal changes or contamination from
sources such as septic tanks, underground fuel storage tanks, or agricultural waste and
chemicals. When more frequent than annual testing is performed by the dialysis facility, testing
may be limited to specific contaminants that are known to be of concern.
Frequency and methods for monitoring water bacteriology are addressed at V252 & V255.
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
9ec3ef99da7f228fb03b74cc1b6934583d83ce4342574ab22cb82e3ca1939493
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.