§494.40 - Condition: Water and dialysate quality.
Interpretive Guidance § 494.40
This Condition incorporates the Association for the Advancement of Medical Instrumentation’s
(AAMI’s) “American National Standard for Dialysate for Hemodialysis,” 2004 (RD52:2004)
into the regulation. This AAMI document references portions of their “American National
Standard for Water Treatment Equipment for Hemodialysis Applications” (RD62:2001) as the
specifications for various water treatment components. The referenced portions of RD62:2001
are also incorporated into the regulation by reference.
The V-tags associated with the water and dialysate quality requirements are divided into 3 parts:
(1) the 2004 AAMI standard, (2) any related 2004 AAMI guidance that CMS identifies as
relevant, and (3) “Additional Guidance” where CMS provides its additional suggestions on how
to best comply with the AAMI guidance.
A survey of this Condition requires inspection of the water treatment and dialysate preparation
equipment, as well as their distribution systems; an interview with personnel responsible for the
day-to-day operation of those systems; and a review of the records of operation and testing for
those systems. Supervisory personnel may be interviewed to clarify issues or questions. Critical
to ensuring patient safety is the expectation that every survey visit include direct observation of
water testing for chlorine/chloramine.
Noncompliance at the Condition level should be considered if identified deficient practices are
pervasive throughout the Standards included in this Condition, serious in nature, and/or a
potential risk to patient health and safety.
Examples of potential Condition level noncompliance may include, but are not limited to:
• Demonstrated lack of knowledge or training of staff assigned responsibility for the
operation or monitoring of the water treatment or dialysate preparation systems;
• Failure to perform and document the test(s) for chlorine and chloramine accurately,
including the use of testing strips or reagents that are expired or not sensitive to the
required levels;
• Unsafe practices in the preparation, labeling, or delivery of dialysate; and,
• Failure to address out-of-range results for tests of water or dialysate (bacteria,
endotoxin, or chemical analysis).
History
Rev.
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
4f5b451728bfbe6b9f3f9f291b62a84271fd9bf404cced2b7b2120f42f65aba1
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.