US · guidance
CMS SOM App. C, Tag D5783
§493.1282 Standard: Corrective actions
(b)(2) Results of control or calibration materials, or both, fail to meet the
laboratory’s established criteria for acceptability. All patient test results obtained
in the unacceptable test run and since the last acceptable test run must be evaluated
to determine if patient test results have been adversely affected. The laboratory
must take the corrective action necessary to ensure the reporting of accurate and
reliable patient test results.
Interpretive Guidelines §493.1282(b)(2)
When an internal control fails to fall within the defined limits of acceptability, the
laboratory must identify the reason for the failure and correct the problem before
resuming testing of patients. The laboratory must evaluate all patient test results since
the last acceptable external control.
Probes §493.1282(b)(2)
When suboptimal staining or improper coverslipping are identified through quality
control procedures, what corrective actions does the laboratory take?
What actions does the laboratory take when controls reflect an unusual trend or are
outside of the acceptable limits and other means of assessing and correcting unacceptable
control values have failed to identify and correct the problem?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
fca0f33af0c7ca95007acf9e085bd26b2728866712e214547e8adf49fdc0330d
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.