CMS SOM App. C, Tag D5525
§493.1264 Standard: Parasitology
(b) The laboratory must calibrate and use the calibrated ocular micrometer for
determining the size of ova and parasites, if size is a critical parameter.
Interpretive Guidelines §493.1264(b)
Check for the following:
• Presence of an ocular micrometer for the microscope(s) used;
• Availability of a stage micrometer;
• Instructions for calibration. Use D5403;
• Records of the measurements and calculations used to show that each objective
(high, oil, low) has been calibrated; and
• Criteria for the use of the micrometer for determining the size of ova and
parasites. Use D5403.
Probes §493.1264(b)
How has the laboratory determined the accuracy of the ocular calibration and that the
staff has the knowledge for proper use?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
4773ecc6fa1b7bc571fc6ef916a8db2703a3a5bfe89d3bdc993ea1d7364c672b
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.