CMS SOM App. C, Tag D5469
§493.1256 Standard: Control procedures
(d)(10) Establish or verify the criteria for acceptability of all control materials.
(d)(10)(i) When control materials providing quantitative results are used, statistical
parameters (for example, mean and standard deviation) for each batch and lot
number of control materials must be defined and available.
(d)(10)(ii) The laboratory may use the stated value of a commercially assayed
control material provided the stated value is for the methodology and
instrumentation employed by the laboratory and is verified by the laboratory.
(d)(10)(iii) Statistical parameters for unassayed control materials must be
established over time by the laboratory through concurrent testing of control
materials having previously determined statistical parameters.
Interpretive Guidelines §493.1256(d)(10)
Acceptable ranges must be verified (assayed) or established (unassayed) by the
laboratory for control materials and any calibrators that are used in lieu of control
materials.
For procedures in which a spiked sample is used as a control, an acceptable range must
be established for the amount of recovery of the spiked sample, either in percentage or
actual concentration.
If laboratories rely on commercial companies to establish statistical limits for controls,
the laboratory must have documentation to verify that its control results correlate with the
established limits.
When patient specimens are used to meet the control requirements, data must be
evaluated in accordance with §493.1256(d)(10)(iii).
There are no specific guidelines for the number of times a material must be tested to
establish statistical limits. In general, twenty replicate tests should be considered the
minimum for determining a standard deviation.
Probes §493.1256(d)(10)
What statistics does the laboratory have to demonstrate the number of assays and the
period of time in which the laboratory repetitively tested control materials to verify or
establish control limits?
How does the laboratory evaluate control results to detect any outliers, shifts or trends in
control values due to instrument malfunctions or changes in the analytical system?
If more than one test system is in use for a test procedure, did the laboratory evaluate the
data for each test method in the establishment of control limits?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
da6116bf761538ab74ab766979ce901c8d1423023698ac4730b3c7bea9b07166
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.