§493.1256 Standard: Control procedures
(d)(6) Perform control material testing as specified in this paragraph before
resuming patient testing when a complete change of reagents is introduced; major
preventive maintenance is performed; or any critical part that may influence test
performance is replaced.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
ff6ed8e0fc5177f2a8d51dc56c381a592c1e196b908f4b2ae47e02a0355d57a9
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.