Bindinglaw

US · guidance

CMS SOM App. C, Tag D5439

§493.1255 Standard: Calibration and calibration verification

activein force · 2026-07-22 – presentas-observed

procedures.

(b) Perform and document calibration verification procedure -

(b)(1) Following the manufacturer’s calibration verification instructions;

(b)(2) Using the criteria verified or established by the laboratory under

§493.1253(b)(3)--

(b)(2)(i) Including the number, type, and concentration of the materials, as well as

acceptable limits for calibration verification; and

(b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a

maximum value near the upper limit of the range to verify the laboratory’s

reportable range of test results for the test system; and

(b)(3) At least once every 6 months and whenever any of the following occur:

(b)(3)(i) A complete change of reagents for a procedure is introduced, unless the

laboratory can demonstrate that changing reagent lot numbers does not affect the

range used to report patient test results, and control values are not adversely

affected by reagent lot number changes.

(b)(3)(ii) There is major preventive maintenance or replacement of critical parts

that may influence test performance.

(b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the

laboratory’s acceptable limits, and other means of assessing and correcting

unacceptable control values fail to identify and correct the problem.

(b)(3)(iv) The laboratory’s established schedule for verifying the reportable range

for patient test results requires more frequent calibration verification.

Interpretive Guidelines §493.1255(b)

The calibration verification requirements may be met by verifying the procedure using a

high-level material such as a control, calibration material, or patient specimen and

diluting it to cover the reportable range if allowed by the manufacturer.

Control activities routinely used to satisfy the requirement for §493.1256 do not satisfy

the calibration verification requirements.

EXCEPTIONS:

1. Laboratories must perform and document calibration procedures following the

manufacturer’s test system instructions, using calibration materials provided or specified,

and at a frequency that is recommended by the manufacturer. Where the manufacturer

does not provide such instruction, the laboratory may calibrate using 3 or more levels of

calibration materials that include a low, mid, and high value at least every 6 months.

2. If the laboratory performs a calibration protocol using 3 or more levels of calibration

materials that include a low, mid, and high value at least every 6 months, the calibration

verification requirement is met.

3. For automated cell counters, the calibration verification requirements are considered

met if the laboratory follows the manufacturer’s instructions for instrument operation and

tests 2 levels of control materials each day of testing provided the control results meet the

laboratory’s criteria for acceptability. This exception does not apply to centrifugal

hematology test systems.

4. For automated chemistry analyzers, the calibration verification requirements are

considered met if the laboratory follows the manufacturer’s instructions for instrument

operation and routinely tests three levels of control materials (lowest level available, mid-level, and highest level available) more than once each day of testing, the control material

results meet the laboratory’s criteria for acceptability and the control materials are

traceable to National Institute of Standards and Technology (NIST) reference materials.

Calibration materials, proficiency testing samples with known results, or control

materials with known values may be used to perform calibration verification. For these

materials, the laboratory must define acceptable limits for the difference between the

measured value obtained, versus the actual concentration of the materials.

NOTE: PT samples can only be used after the event cut-off date.

“Calibration material” means a solution that has a known amount of analyte weighed in,

has a value determined by repetitive testing using a reference/definitive test method or is

traceable to National Institute of Standards and Technology (NIST) reference material, if

possible.

If a manufacturer provides reagents for a test where all of the reagents for a test are

packaged together, calibration verification is not required for each additional reagent

package with the same lot number that is received in the same shipment. For example, if

the laboratory receives 12 packs of reagents and the laboratory has verified calibration for

at least one of the 12 packs of reagents, then the laboratory does not have to verify

calibration for the remaining 11 packs of reagents provided that all 12 packs of reagents

have the same lot number and were received on the same shipment to the laboratory.

However, this exception does not override the requirement to perform calibration

verification as specified at 493.1255(b)(3).

5. Calibration verification is not required on:

• Instruments that are factory or manufacturer calibrated and/or

• Tests that are considered non-quantitative (e.g., Prothrombin time and Activated

Clotting Time, which are measured in units of time)

When reviewing the laboratory’s maintenance and function check records as required in

§493.1254, determine whether the laboratory performed calibration verification when

major maintenance occurred or critical parts were replaced.

The actual measurement(s) taken, reactions and/or observations must be recorded.

Probes §493.1255(b)

If a laboratory does not perform calibration verification after a complete change of

reagents, what data does the laboratory have to document that changing reagent lot

numbers does not affect the reportable range of patient test results, and does not

adversely affect control results?

History

Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
e2bbe9b725b9c09663ddb19b7172e2bfd798cd504a851864be2c3847c51f3ea9
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.