US · guidance
CMS SOM App. C, Tag D5437
§493.1255 Standard: Calibration and calibration verification
procedures.
(a) Perform and document calibration procedures -
(a)(1) Following the manufacturer’s test system instructions, using calibration
materials provided or specified, and with at least the frequency recommended by
the manufacturer;
(a)(2) Using the criteria verified or established by the laboratory as specified in
§493.1253(b)(3)--
(a)(2)(i) Using calibration materials appropriate for the test system and, if possible,
traceable to a reference method or reference material of known value; and
(a)(2)(ii) Including the number, type, and concentration of calibration materials, as
well as acceptable limits for and the frequency of calibration; and
(a)(3) Whenever calibration verification fails to meet the laboratory’s acceptable
limits for calibration verification.
Interpretive Guidelines §493.1255(a)
Laboratories must follow the manufacturer’s instructions on carrying out the calibration
and must follow or exceed the manufacturer’s frequency recommendations for
calibration. However, if a calibration system proves less stable than expected by the
manufacturer, additional calibration materials and/or more frequent calibration may be
required, as established or verified by the laboratory under §493.1253(b)(3).
The calibration requirement does not apply to a variety of procedures, which include, but
are not limited to:
• Manual procedures not involving an instrument (e.g., microbiology cultures,
Kirby-Bauer disk susceptibility tests, tilt-tube prothrombin time test systems,
ABO group and D (Rho) typing);
• Microscopic procedures (e.g., KOH preparations, pinworm preparations, urine
sediment analysis, all manual differential procedures, manual cytology screening
procedures); and
• Test systems which include instruments that cannot be adjusted or calibrated
because they are factory or manufacturer calibrated (e.g. unit use devices). This
would include prothrombin time procedures on a fibrometer, or instruments that
utilize a whole blood specimen and single unit use cartridge (PT/INR, Activated
Clotting Time).
The term “calibration material” has generally replaced “standard” since many instruments
now use serum-based reference materials. “Calibration material” means a solution that
has a known amount of analyte weighed in or has a value determined by repetitive testing
using a reference/definitive test method or is traceable to a National Institute for
Standards and Technology (NIST) Standard, if possible.
The actual measurement(s) taken, reactions, and/or observations must be recorded.
Probes §493.1255(a)
If the laboratory calculates values for one or more calibration materials, are the
calculations correct, and do the records reflect that the measured values are within the
laboratory’s established limits for the calibration materials?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
82907c609040bb214bffdd14f917d7fe69c36971f04efb70d9e94e45b8641806
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.