Bindinglaw

US · guidance

CMS SOM App. C, Tag D5435

§493.1254 Standard: Maintenance and function checks

activein force · 2026-07-22 – presentas-observed

(b)(2)(i) Define a function check protocol that ensures equipment, instrument, and

test system performance that is necessary for accurate and reliable test results and

test result reporting.

(b)(2)(ii) Perform and document the function checks, including background or

baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks

must be within the laboratory’s established limits before patient testing is

conducted.

Interpretive Guidelines §493.1254(b)(2)(i)-(b)(2)(ii)

The laboratory must establish and follow procedures for performing function checks on

each piece of equipment/instrument it uses, including those that are peripherally involved

in patient testing (e.g., incubators, centrifuges, safety cabinets, autoclaves).

Function checks refer to those activities performed to evaluate critical operating

characteristics (e.g., stray light, zeroing, electrical levels, optical alignment, background

counts, counting efficiency) according to the accepted method of operation for each type

of device or instrument. Daily quality control activities and function checks are

performed prior to patient testing to ensure that an instrument is functioning correctly and

is properly calibrated. Checking electrical, mechanical, and operational functions may be

independent of the procedure. The performance of daily quality control activities serves

as an additional instrument function check. Analysis of external control samples check

the operating characteristics of a test system, including instrument stability and

calibration.

When function checks are critical to test performance, the laboratory must have a

mechanism in place to monitor such items as:

• Rotator speed and circumference;

• Timers;

• Anaerobic chambers;

• Cell washers;

• Radioactive particle counters;

• Blood cell counters; and

• Nucleic acid amplification equipment.

Flow Cytometry:

A fluorescence standard(s) for each fluorochrome must be used each day of patient

testing to ensure:

• Proper alignment of the optical system;

• Standardization of the fluorescence detectors;

• Resolution of dimly-stained particles; and

• Appropriate compensation for spectral overlap of the fluorochromes.

Fluorescence standards must have the same fluorochromes incorporated into them as are

used for the test, and with the exception of alignment standards, must have similar

fluorescence intensities as found in the test specimens. The laboratory must have an

acceptable range of performance for all procedures.

For flow cytometers with air-cooled lasers, the laser should be tested each day patients

are tested by peaking the laser signal and monitoring the current input (amps) to laser

light output (milliwatts) to determine whether the brewster windows are in need of

cleaning.

Probes §493.1254(b)(2)

For those methods in which the centrifugation is a critical portion of the test, how has the

laboratory checked the established RPM’s and timing as necessary?

In immunofluorescent test procedures, how does the laboratory ensure that the bulb is

emitting ultraviolet light at the correct wavelength?

If function checks are not required or recommended by the manufacturer, how does the

laboratory establish the performance criteria of its equipment and instruments?

For RIA testing, are backgrounds or baselines measured for each setting? For example, if

the laboratory uses more than one type of isotope, at what window setting are background

counts performed and recorded?

When performing flow cytometry analysis using two or more fluorochromes

simultaneously, how does the laboratory identify and adjust for “spill over” into the other

fluorescence detectors?

§493.1255 Standard: Calibration and calibration verification

procedures.

Calibration and calibration verification procedures are required to substantiate the

continued accuracy of the test system throughout the laboratory’s reportable range

of test results for the test system. Unless otherwise specified in this subpart, for each

applicable test system the laboratory must do the following:

Interpretive Guidelines §493.1255

For definitions of calibration and calibration verification, refer to §493.2.

For calibration and calibration verification of blood gas analysis, see §493.1267(a)

through (d).

In many instances, the performance of method calibration serves to satisfy the

requirement for instrument calibration. Calibration procedures are not to be confused

with instrument/equipment function checks at §493.1254.

History

Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25

Provenance

Source
cms.gov
Retrieved
2026-07-22
Edition
som-2026-07-22
Content hash
11b97f801b2a4592f0b880dcb8e5d315426fac2e94e0966646b365489c17b7dd
View the official source →

The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.

Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.

Coverage · API docs

Bindinglaw

Point-in-time US law with the receipt attached. Source URL, retrieval time, content hash, and validity dates on every answer.

curl api.binding.law/v1/law/coverage

© 2026 binding.law · a Jubal, Inc. productAttorneys and firms never pay. Ever.