US · guidance
CMS SOM App. C, Tag D5435
§493.1254 Standard: Maintenance and function checks
(b)(2)(i) Define a function check protocol that ensures equipment, instrument, and
test system performance that is necessary for accurate and reliable test results and
test result reporting.
(b)(2)(ii) Perform and document the function checks, including background or
baseline checks, specified in paragraph (b)(2)(i) of this section. Function checks
must be within the laboratory’s established limits before patient testing is
conducted.
Interpretive Guidelines §493.1254(b)(2)(i)-(b)(2)(ii)
The laboratory must establish and follow procedures for performing function checks on
each piece of equipment/instrument it uses, including those that are peripherally involved
in patient testing (e.g., incubators, centrifuges, safety cabinets, autoclaves).
Function checks refer to those activities performed to evaluate critical operating
characteristics (e.g., stray light, zeroing, electrical levels, optical alignment, background
counts, counting efficiency) according to the accepted method of operation for each type
of device or instrument. Daily quality control activities and function checks are
performed prior to patient testing to ensure that an instrument is functioning correctly and
is properly calibrated. Checking electrical, mechanical, and operational functions may be
independent of the procedure. The performance of daily quality control activities serves
as an additional instrument function check. Analysis of external control samples check
the operating characteristics of a test system, including instrument stability and
calibration.
When function checks are critical to test performance, the laboratory must have a
mechanism in place to monitor such items as:
• Rotator speed and circumference;
• Timers;
• Anaerobic chambers;
• Cell washers;
• Radioactive particle counters;
• Blood cell counters; and
• Nucleic acid amplification equipment.
Flow Cytometry:
A fluorescence standard(s) for each fluorochrome must be used each day of patient
testing to ensure:
• Proper alignment of the optical system;
• Standardization of the fluorescence detectors;
• Resolution of dimly-stained particles; and
• Appropriate compensation for spectral overlap of the fluorochromes.
Fluorescence standards must have the same fluorochromes incorporated into them as are
used for the test, and with the exception of alignment standards, must have similar
fluorescence intensities as found in the test specimens. The laboratory must have an
acceptable range of performance for all procedures.
For flow cytometers with air-cooled lasers, the laser should be tested each day patients
are tested by peaking the laser signal and monitoring the current input (amps) to laser
light output (milliwatts) to determine whether the brewster windows are in need of
cleaning.
Probes §493.1254(b)(2)
For those methods in which the centrifugation is a critical portion of the test, how has the
laboratory checked the established RPM’s and timing as necessary?
In immunofluorescent test procedures, how does the laboratory ensure that the bulb is
emitting ultraviolet light at the correct wavelength?
If function checks are not required or recommended by the manufacturer, how does the
laboratory establish the performance criteria of its equipment and instruments?
For RIA testing, are backgrounds or baselines measured for each setting? For example, if
the laboratory uses more than one type of isotope, at what window setting are background
counts performed and recorded?
When performing flow cytometry analysis using two or more fluorochromes
simultaneously, how does the laboratory identify and adjust for “spill over” into the other
fluorescence detectors?
§493.1255 Standard: Calibration and calibration verification
procedures.
Calibration and calibration verification procedures are required to substantiate the
continued accuracy of the test system throughout the laboratory’s reportable range
of test results for the test system. Unless otherwise specified in this subpart, for each
applicable test system the laboratory must do the following:
Interpretive Guidelines §493.1255
For definitions of calibration and calibration verification, refer to §493.2.
For calibration and calibration verification of blood gas analysis, see §493.1267(a)
through (d).
In many instances, the performance of method calibration serves to satisfy the
requirement for instrument calibration. Calibration procedures are not to be confused
with instrument/equipment function checks at §493.1254.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
11b97f801b2a4592f0b880dcb8e5d315426fac2e94e0966646b365489c17b7dd
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.