US · guidance
CMS SOM App. C, Tag D5431
§493.1254 Standard: Maintenance and function checks
(a)(2) Function checks as defined by the manufacturer and with at least the
frequency specified by the manufacturer. Function checks must be within the
manufacturer’s established limits before patient testing is conducted.
Interpretive Guidelines §493.1254(a)(2)
Function checks refer to those activities performed to evaluate critical operating
characteristics (e.g., stray light, zeroing, electrical levels, optical alignment, background
counts, counting efficiency) according to the accepted method of operation for each type
of device or instrument. Daily quality control activities and function checks are
performed prior to patient testing to ensure that an instrument is functioning correctly and
is properly calibrated (Checking electrical, mechanical, and operational functions may be
independent of the procedure). The performance of daily quality control activities may
serve as an additional instrument function check, since analysis of external control
samples check the operating characteristics of a test system, including instrument stability
and calibration.
The laboratory must follow and document the required functions checks as stated by the
laboratory information system (LIS) manufacturer for the LIS computer and devices such
as monitors, printers and modems.
For instruments that automatically perform function checks and flag problems, the
laboratory is required to document the corrective actions in response to the flagged
problems. Use D5793 for deficiencies related to documenting corrective actions in
response to the flagged problems.
Flow Cytometry:
A fluorescence standard(s) for each fluorochrome should be used each day of patient
testing to ensure:
• Proper alignment of the optical system;
• Standardization of the fluorescence detectors;
• Resolution of dimly-stained particles; and
• Appropriate compensation for spectral overlap of the fluorochromes.
Fluorescence standards should have the same fluorochromes as are used for the test, and
with the exception of alignment standards, should have similar fluorescence intensities as
found in the test specimens. The laboratory must have an acceptable range of
performance for all procedures.
Probes §493.1254(a)(2)
For those methods in which the centrifugation is a critical portion of the test, does the
laboratory check the RPM’s and timing periodically (e.g., urine sediments)?
Do the records of a laboratory that moves from testing site to testing site demonstrate the
performance of function checks as necessary?
In immunofluorescent test procedures, how does the laboratory ensure that the bulb is
emitting ultraviolet light at the correct wavelength?
How does the laboratory ensure that the fluorescent light source has not exceeded the
manufacturer’s established optimal timeframe?
For procedures or test systems that require pipetting or dilution of patient specimens
separately from controls or calibrators, how are autodiluters, microdiluters, and/or
pipettors checked for adequate and consistent delivery?
For those systems that perform simultaneous fluid delivery to multi-well plates or tubes,
how does the laboratory ensure uniform delivery of reagents or washing solutions to all
wells or tubes?
§493.1254 Standard: Maintenance and function checks.
(b) Equipment, instruments, or test systems developed in-house, commercially
available and modified by the laboratory, or maintenance and function check
protocols are not provided by the manufacturer. The laboratory must do the
following:
Interpretive Guidelines §493.1254(b)
The laboratory must establish and follow procedures for performing maintenance and
function checks on each piece of equipment/instrument it uses, including those that are
peripherally involved in patient testing (e.g., incubators, centrifuges, safety cabinets,
autoclaves and microscopes).
A manufacturer’s instructions may not require maintenance and function checks.
However, if the laboratory determines that a maintenance and/or function check protocol
is necessary in order to ensure accurate and reliable test results, the laboratory must
establish a maintenance and/or function check protocol and perform and document the
described activities as they are carried out over time.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
cebb43e297e0a52e086ed7308fdb6619356c11a3a55a2b9b365977b206312629
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.