CMS SOM App. C, Tag D5425
§493.1253 Standard: Establishment and verification of performance
specifications.
(b)(3) Determination of calibration and control procedures. The laboratory must
determine the test system’s calibration procedures and control procedures based
upon the performance specifications verified or established under paragraph (b)(1)
or (b)(2) of this section.
Interpretive Guidelines §493.1253(b)(3)
Through the verification/establishment process, the laboratory defines the frequency for
calibration and control performance as well as the type, number, and concentration of
calibration and control materials used to monitor, detect error, and evaluate method
performance. The frequency for calibration and control performance must not be less
than the frequency specified in the manufacturer’s instructions.
In establishing the calibration and quality control frequency, the laboratory must
consider:
• Test system instrument/reagent stability, including relocation;
• Frequency with which the test is performed;
• Technique dependence of the method;
• Frequency of quality control failures; and
• Training, experience, and competency of technical personnel.
For additional criteria in determining calibration and quality control frequency refer to
§§493.1255 and 493.1256.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
1b0656554bb2c1e93432907338d4d960542d566227836536a5dc5e134df721c4
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.