US · guidance
CMS SOM App. C, Tag D5421
§493.1253 Standard: Establishment and verification of performance
specifications.
(b)(1) Verification of performance specifications. Each laboratory that introduces
an unmodified, FDA-cleared or approved test system must do the following before
reporting patient test results:
Interpretive Guidelines §493.1253(b)(1)
The laboratory is responsible for verifying the performance specifications of each
nonwaived unmodified FDA-cleared or approved test system that it introduces, prior to
reporting patient test results. The verification of method performance should provide
evidence that the accuracy, precision, and reportable range of the procedure are adequate
to meet the clients’ needs, as determined by the laboratory director and clinical
consultant. A laboratory may use the manufacturer’s performance specifications as a
guideline, but is responsible for verifying the manufacturer’s analytical claims before
initiating patient testing. Refer to CLIA Brochure - Verification of Performance
Specifications (PDF).
If a method was verified by someone other than the laboratory staff (e.g., manufacturer
representative), the laboratory must demonstrate that this verification correlates with its
in-house test performance. This may be accomplished by the laboratory testing “known”
samples.
For some qualitative tests, the laboratory may verify the manufacturer’s specifications by
testing known positive and negative samples to ensure that the expected results are
obtained. (Specimens of known quantitative value may be used to verify the accuracy of
a qualitative test.)
Prior to introducing a test for routine patient testing, the laboratory must review and
evaluate the verification data.
Each laboratory is responsible for determining that its performance specifications for
each test system are not affected by the relocation of the laboratory or test system. (See
manufacturer’s package insert regarding critical requirements such as set-up, limitations,
environmental conditions, etc.) When a temporary replacement (loaner) instrument is
received which is identical (i.e., same make and model, and method for the same analyte)
to the instrument which is being replaced, the laboratory must verify performance
specifications.
If calibration material is used to verify method performance specifications, the laboratory
must demonstrate that there is a minimal matrix effect and the calibration material is
appropriate for verifying test system performance specifications.
If the LIS performs any calculations to determine a laboratory result, the calculations
must be verified immediately after the LIS is programmed and prior to initial calculation
of patient results.
“Less than” is used for reporting test results that are below the laboratory’s detection
limits for an analyte. (Detection limits must be established through method verification.)
“Equivalent designation” is used to report test results for those methods that yield results
below a clinically significant level (e.g., for a quantitative immunology test, patient
results may be clinically negative at a 1:8 titer and test results may be reported as “1:8
negative”). (The normal value is 1:8 or less.) “Greater than” is used for reporting test
results that are above the laboratory’s detection limits for an analyte. If patient test
results exceed the laboratory’s reportable range, the laboratory must report the result as
greater than the highest detection limit, re-assay a diluted patient specimen and report the
calculated result, or send the specimen to a reference laboratory.
CDC Developed Assays and Non-CDC Developed Assays:
Laboratories are required to verify performance specifications for CDC developed EUA
assays per 42 CFR 493.1253(b)(1). In general, CDC developed test kits provide an initial
set of samples for verifying performance specifications and instructions for their use.
Laboratories using a CDC developed assay authorized for emergency use should follow
any and all instructions provided for verifying performance specifications.
As with CDC developed EUA assays, laboratories are required to verify performance
specifications for other, non-CDC developed EUA assays per 42 CFR 493.1253(b)(1).
Laboratories are required to verify performance specifications according to the
performance specifications established by the manufacturer and the manufacturer’s
reference intervals. In accordance with 42 CFR §493.1252(a), any manufacturer’s
instructions for verification must be followed, if provided. The surveyor should confirm
that the laboratory has followed its procedures for verification.
Refer to CMS QSO-18-19-CLIA, Performance Specification Verification of Assays
Authorized Under Emergency Use (EUA) for additional information.
Probes §493.1253(b)(1)
How does the laboratory determine if a new or revised LIS program (whether purchased
or developed in-house) performs acceptably before it is integrated into routine operation?
(b)(1)(i) Demonstrate that it can obtain performance specifications comparable to
those established by the manufacturer for the following performance
characteristics:
Interpretive Guidelines §493.1253(b)(1)(i)
Laboratories may simultaneously verify multiple performance specifications by choosing
appropriate samples; e.g., repeatedly test (precision) samples with known (accuracy) high
and low values (reportable range). This testing should be performed among all operators
on different days. In addition, for test systems of the same make and model, consider
verifying performance specifications of these devices at the same time.
§493.1253 Standard: Establishment and verification of performance
specifications.
(b)(1)(i)(A) Accuracy.
Interpretive Guidelines §493.1253(b)(1)(i)(A)
Accuracy- The laboratory is responsible for verifying that the method produces correct
results. Verification of accuracy may be accomplished by:
• Testing reference materials;
• Comparing results of tests performed by the laboratory against the results of a
reference method; or
• Comparing split sample results with results obtained from another method, which
has already been shown to provide accurate results.
For qualitative methods, the laboratory must verify that a method will identify the
presence/absence of the analyte.
§493.1253 Standard: Establishment and verification of performance
specifications.
(b)(1)(i)(B) Precision.
Interpretive Guidelines §493.1253(b)(1)(i)(B)
Precision (Reproducibility) - The laboratory is responsible for verifying the precision of
each test system by assessing day-to-day, run-to-run, and within-run variation, as well as
operator variance. This may be accomplished by:
• Repeat testing of known patient samples over time;
• Testing QC material in duplicate and over time; or
• Repeat testing of calibration materials over time.
EXCEPTION: For fully automated systems that are not user dependent, operator
variance does not need to be evaluated.
(b)(1)(i)(C) Reportable range of test results for the test system.
Interpretive Guidelines §493.1253(b)(1)(i)(C)
Reportable Range- The laboratory is responsible for verifying the reportable range of
patient test results for each test system. Verification of reportable range may be
accomplished by:
• Assaying low and high calibration materials or control materials; or
• Evaluating known samples of abnormal high and abnormal low values.
Hematology whole blood high range calibration materials are not generally available.
Therefore, laboratories may use patient specimens with verified elevated cell counts to
verify the upper limit of the reportable range.
Probes §493.1253(b)(1)(i)(C)
If a dilution procedure is used when patient results exceed the test system’s reportable
range, how does the laboratory ensure the appropriate diluent is used for each type of
specimen?
How does the laboratory verify and document the accuracy of the results for diluted
specimens?
§493.1253 Standard: Establishment and verification of performance
specifications.
(b)(1)(ii) Verify that the manufacturer’s reference intervals (normal values) are
appropriate for the laboratory’s patient population.
Interpretive Guidelines §493.1253(b)(1)(ii)
Reference Range (Normal Values) - The laboratory may use the manufacturer’s reference
range provided it is appropriate for the laboratory’s patient population (i.e., a normal
range that reflects the type of specimen and demographic variables such as age and sex,
as applicable). If the manufacturer has not provided reference ranges appropriate for the
laboratory’s patient population, the laboratory may use published reference range(s). The
laboratory must evaluate an appropriate number of specimens to verify the
manufacturer’s claims for normal values or, as applicable, the published reference ranges.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
53a385745502232844cee439ced71241e02ae21d369dfd3fa748a44f9fc432e4
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.