(b) The laboratory must have appropriate and sufficient equipment, instruments,
reagents, materials, and supplies for the type and volume of testing it performs.
Interpretive Guidelines §493.1101(b)
Base deficiencies related to inappropriate or insufficient equipment on a determination
that patient results are or may be adversely affected. Ensure that the laboratory has the
appropriate equipment to prepare reagents, stains, solutions, controls, and calibration
materials (e.g., pipettes, hydrometers, graduated cylinders, autoclaves, balances,
centrifuges, distilled/deionized water). If the equipment or instrumentation is found to be
inappropriate or insufficient, document the reasons for this finding.
Ensure that the laboratory has test systems, equipment and/or instruments capable of
producing results within the laboratory’s stated test performance specifications.
Ensure that the laboratory has test systems, equipment and/or instruments necessary to
perform the laboratory’s volume of testing (preanalytic, analytic, postanalytic) within
established turnaround times.
Data capacity in the laboratory’s information system should be sufficient for current data
entry. If capacity is maintained by deletion of data, it should be scheduled and
documented.
For Cytology, laboratories should use coverslips that cover the entire surface of the
specimen.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
e54f8410b349b866456150411e74ab2da4621df6e68e666fa52cb02e719846a9
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.