§493.1101 Standard: Facilities
(a)(3) Molecular amplification procedures that are not contained in closed systems
have a uni-directional workflow. This must include separate areas for specimen
preparation, amplification and product detection, and, as applicable, reagent
preparation.
Interpretive Guidelines §§493.1101(a)(2)-(a)(3)
The laboratory should establish contamination prevention procedures to minimize
contamination of patient specimens, equipment, instruments, reagents, materials, and
supplies.
Determine if the laboratory performs wipe tests of areas where radioactive material or
amplification procedures are used in order to monitor and prevent contamination.
Laboratories performing molecular amplification procedures should have a mechanism to
detect cross-contamination of patient specimens. This may be accomplished by including
a “blank” control with each run of patient specimen testing (use D5425).
The “blank” control refers to a no-template control (N.T.C) or a control sample
containing all reagents except the target template.
An example of a “closed system” would be an FDA-cleared or FDA-approved test
system that contains amplification and detection steps in sealed tubes that are never
opened or re-opened during or after the testing process and that is used as directed or
suggested by the manufacturer (i.e., without any modifications).
Unidirectional workflow refers to the manner in which testing personnel and patient
specimens move through the molecular testing process to prevent cross-contamination,
and consists of separate areas for the following:
• Reagent preparation (as applicable);
• Pre-amplification area for specimen preparation and amplification reaction set up;
and
• Post-amplification area for specimen amplification, product detection, and storage
or disposal of amplified products.
Reagents must be prepared in an area that is separate (as applicable) from where
specimens are processed, prepared, “amplified” and detected to prevent contamination.
Once a specimen enters the amplification and product detection area it should not be
brought back to the reagent or specimen preparation areas. The laboratory should store
amplified specimens separately from test reagents and patient specimens. All equipment
(e.g., reagents, supplies, pens, pipettes and tips, laboratory coats) should remain in
designated areas.
Sources of potential cross-contamination in molecular testing include:
• Patient specimen (i.e., genomic contamination);
• Amplified patient specimen (i.e., amplicon contamination); and
• Testing personnel.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
a7ee31a17787b734f09943a5d7df550d34c3d85261bb78917aab438bfbff0c54
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.