US · guidance
CMS SOM App. C, Tag D2016
§493.803 Condition: Successful participation
(a) Each laboratory performing nonwaived testing must successfully participate in a
proficiency testing program approved by CMS, if applicable, as described in
subpart I of this part for each specialty, subspecialty, and analyte or test in which
the laboratory is certified under CLIA.
(b) Except as specified in paragraph (c) of this section, if a laboratory fails to
participate successfully in proficiency testing for a given specialty, subspecialty,
analyte or test, as defined in this section, or fails to take remedial action when an
individual fails gynecologic cytology, CMS imposes sanctions, as specified in subpart
R of this part.
(c) If a laboratory fails to perform successfully in a CMS-approved proficiency
testing program, for the initial unsuccessful performance, CMS may direct the
laboratory to undertake training of its personnel or to obtain technical assistance, or
both, rather than imposing alternative or principle sanctions except when one or
more of the following conditions exists:
(1) There is immediate jeopardy to patient health and safety.
(2) The laboratory fails to provide CMS or a CMS agent with satisfactory
evidence that it has taken steps to correct the problem identified by the
unsuccessful proficiency testing performance.
(3) The laboratory has a poor compliance history.
Interpretive Guidelines §493.803
Only the PT program has the capability to correct scores in the CMS PT monitoring
system. See SOM section 6056.
No single PT enforcement protocol is universally applicable for all situations. Unique
circumstances may require special considerations or actions that may not conform to the
general approach outlined below. The laboratory’s compliance history, its willingness to
take remedial actions, and the professional judgment of surveyors and enforcement
personnel may be factors in determining an appropriate PT enforcement plan.
Careful review of PT performance reports and other available information should always
be performed to determine whether the PT results truly represent failed PT. The potential
of a PT program data input error or other factors beyond the laboratory’s control should
be considered. If the laboratory has made a transcription error(s), it is considered
erroneous PT result(s).
If review and verification of PT performance reports confirm unsuccessful PT, cite as a
Condition-level deficiency (use D2016 on the Form CMS-2567).
NOTE: The CMS PT monitoring system may NOT be used alone to determine
unsuccessful participation. Surveyors must verify any unsuccessful participation
indicated in the PT monitoring system. This may be done by reviewing PT results
supplied by the approved PT program (they will send copies to the surveyor if requested)
or from results sent to the laboratory by the PT program.
If the unsuccessful PT participation is the first occurrence for the laboratory, and none of
the exceptions listed at §493.803(c)(1)-(3) exist, cite the laboratory at the condition-level
(D2016). Notify the laboratory to seek training of its personnel, obtain the necessary
technical assistance to correct the problem causing the unsuccessful participation, or
both. SAs may initiate training and/or technical assistance after first obtaining CMS’
concurrence. No on-site review is required to initiate this action.
The laboratory must submit an allegation of compliance, listing completion dates and
other pertinent information for its training and/or technical assistance efforts. Follow-up
is necessary to verify that the laboratory has carried out its allegation of compliance. If a
laboratory refuses to take acceptable training and/or technical assistance actions
(including failure to submit an allegation of compliance, or failure to complete its
allegation of compliance), a sanction action may be initiated by CMS.
When the unsuccessful PT participation is not the first such occurrence for the laboratory,
cite as a condition-level deficiency and take appropriate enforcement action. In some
cases, it may also be necessary to cite the laboratory director in order for the laboratory to
achieve compliance. Citing of the laboratory director in addition to PT citations, is at the
discretion of the surveyor. For immediate jeopardy cases, the procedures in Subpart R
apply. For non-immediate jeopardy situations, enforcement procedures should be
completed within 90 days from the date that the unsuccessful PT was first identified. In
immediate jeopardy situations, enforcement procedures should be completed within 23
days from the date unsuccessful participation of PT is first identified.
Example 1: Initial Unsuccessful
A laboratory scores 60% on a testing event in Sodium. On the next testing event, the
laboratory fails to participate in Sodium (Na). The citations are D2096 (§493.841(f)) and
D2016 (§493.803). D6000 (§493.1403) or D6076 (§493.1441) are laboratory director
citations. (Note: It is not necessary to cite the standard for unsatisfactory analyte
performance. However, it is necessary to cite the standard when the laboratory fails to
participate in a testing event so that the laboratory is made aware that such deficient
practice results in a score of 0 for the testing event.)
Analyte 2019 2020 2021 2022
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Sodium
(Na)
100%
80%
100%
100%
80%
80%
100%
80%
80%
60%
0%
100%
Example 2: Initial Unsuccessful
A laboratory scores 60% on uric acid PT samples. On the next testing event, the
laboratory scores 40% on the same analyte. The citations are D2096 (§493.841(f)) and
D2016 (§493.803). D6000 (§493.1403) or D6076 (§493.1441) are laboratory director
citations. (Note: Cite the standard for unsuccessful performance and the condition for
unsuccessful participation. It is not necessary to cite the standard for unsatisfactory
analyte performance.)
Analyte 2019 2020 2021 2022
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Uric
Acid
(UA)
100%
100%
80%
100%
100%
100%
80%
80%
100%
60%
40%
100%
Example 3: Non-initial (Subsequent) Unsuccessful - Rolling Timeframe
A laboratory scores 60% in Sodium for two consecutive PT events resulting in initial
unsuccessful PT performance. The citations are D2096 (§493.841(f)) and D2016
(§493.803). The laboratory scores 80% for the next three PT events. After the three
successful PT events, the laboratory scores 60% in two consecutive PT events resulting in
non-initial unsuccessful PT performance. The citations are D2096 (§493.841(f)) and
D2016 (§493.803). D6000 (§493.1403) or D6076 (§493.1441) are laboratory director
citations. Non-initial unsuccessful PT performance is referred to CMS for enforcement
actions. See SOM sections 6050 and 6054.
Analyte 2019 2020 2021 2022
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Event
1
Score
Event
2
Score
Event
3
Score
Sodium
(Na)
100%
80%
100%
80%
100%
60%
60%
80%
80%
80%
60%
60%
When recommending to CMS that a laboratory be subject to sanctions, submit copies of
the laboratory’s testing event or analyte score(s) that were unsatisfactory and the correct
responses provided by the PT program. Also, enclose copies of any correspondence sent
to or received by the laboratory concerning its PT performance.
Probes:
Do the laboratory’s PT results show satisfactory performance in the next PT event?
Did the laboratory verify its corrective action or training and technical assistance was
successful?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
e9b5aca3e5c59de06a82af9756349aaec01a5a49a894a70a764a83748fe58f62
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.