US · guidance
CMS SOM App. C, Tag D2006
§493.801(b) Standard: Testing of proficiency testing samples
(b)The laboratory must examine or test, as applicable, the proficiency testing
samples it receives from the proficiency testing program in the same manner as it
tests patient specimens. This testing must be conducted in conformance with
paragraph (b)(4) of this section. If the laboratory’s patient specimen testing
procedures would normally require reflex, distributive, or confirmatory testing at
another laboratory, the laboratory should test the proficiency testing sample as it
would a patient specimen up until the point it would refer a patient specimen to a
second laboratory for any form of further testing.
Interpretive Guidelines §493.801(b)
Review testing records to determine if special handling was given to PT samples.
Consider the unique requirements of many PT samples when evaluating “same manner”
of testing. The laboratory should document any necessary reconstitution, longer mixing
times, unit conversion of results, etc., as required in §493.801(b)(5).
A laboratory that routinely performs only presumptive testing or screening methods and
refers patient samples to another laboratory for definitive or confirmatory testing or
comparison of test results must not refer PT samples to another laboratory for
confirmatory testing. A laboratory should limit the testing of PT specimens to the
patient testing which is done in-house. With the exception of specimen preparation such
as Immunohistochemistry (IHC) staining, laboratories need to take great care to avoid
sending PT specimens or results to any entity other than their PT provider prior to the PT
testing event cutoff date.
A central laboratory with more than one instrument or methodology for the same test may
alternate methods or instruments from one testing event to the next as long as both are
routinely used to test patient specimens. All samples for one analyte within a shipment
must be tested with the same instrument.
Probes §493.801(b)
• What procedure or test method was used?
• Is this a routine test method used in the laboratory?
• Did routine personnel perform the PT?
• How often were PT samples tested? Does this conform with the laboratory’s
written policies for patient specimens?
• How are deviations from general laboratory practices (if any) justified?
• Do the PT results documented in the laboratory work records (worksheet)
correlate with the results reported to the PT program?
• Do reports submitted to the PT program provider accurately reflect the procedure
(i.e., instrument, method) used in the laboratory?
Check to see if patient samples were reported on the same day that PT samples were
tested. (In a small facility, infrequent testing may necessitate the testing of PT samples
without patient specimens to ensure that the PT test results are returned on time.) Did the
laboratory use the same procedure for both patient specimens and PT samples?
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
d6d1f08a69d4be6078061ced5dd76487da0bc5d3f667f9c14e54f64e8cb975f2
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.