US · guidance
CMS SOM App. C, Tag D1001
§493.15(e) Laboratories eligible for a certificate of waiver must--
(1) Follow manufacturers’ instructions for performing the test; and
(2) Meet the requirements in subpart B, Certificate of Waiver, of this part.
Interpretive Guidelines §493.15(e)
Tests listed on the waiver list in §493.15(c) are not subject to routine survey. A survey
of waived tests may be conducted only when authorized by CMS in the following
instances:
• Determine if a laboratory is testing outside its certificate;
• Collect information regarding the appropriateness of tests specified as waived
tests;
• Investigate a complaint from the public; and/or
• Determine if the laboratory is operated and if testing is performed in a manner
that does not constitute an imminent and serious risk to public health.
Refer to §§493.1773 and 493.1775 for additional guidelines for inspecting laboratories
issued a certificate of waiver.
Laboratories holding a certificate of waiver must follow the current manufacturer’s
instructions for the waived test systems they are using for patient testing. To meet the
waived testing regulatory requirements, these laboratories must comply with the
manufacturer’s requirements. We encourage laboratories to also comply with the
manufacturer’s recommendations for testing. These laboratories may only use the
specimen types that were approved by the Food and Drug Administration (FDA) with the
waived test system they are using, and they must follow the manufacturer’s quality
control (QC) and test performance requirements. We encourage laboratories to also
comply with manufacturer’s recommendations for the waived test system. Some
manufacturers produce tests that can be run as a waived test or a moderate complexity
test. Any laboratory with a certificate of waiver that uses the nonwaived test system
instructions from a manufacturer should be advised that they must use the manufacturer’s
instructions for waived testing. If the situation remains uncorrected, the laboratory may
be cited for performing tests beyond the scope of the certificate held by the laboratory, as
well as failing to follow manufacturer’s instructions.
NOTE: It is never acceptable for a laboratory operating under a certificate of waiver to
modify the manufacturer’s instructions for the waived test system. Any such changes
will result in a test that is no longer waived (i.e., the waived test is uncategorized for
CLIA and therefore becomes a high complexity test). For example, if a test specifies
urine as the waived specimen type and the laboratory tests a different body fluid, then the
laboratory is no longer performing a waived test and the lab is then subject to routine
inspections and the CLIA requirements for high complexity testing. Waived laboratory
testing personnel must follow the manufacturer’s instructions in their entirety and without
variation. Great care should be taken to add the proper reagents in the order and amount
specified by the manufacturer’s instructions to ensure compliance with the CLIA
regulations and reliable test results.
§493.17 Test categorization
(Rev. 140, Issued: 05-29-15, Effective: 05-29-15, Implementation: 05-29-15)
(a) Categorization by criteria. Notices will be published in the FEDERAL
REGISTER which list each specific test system, assay, and examination categorized
by complexity. Using the seven criteria specified in this paragraph for categorizing
tests of moderate or high complexity, each specific laboratory test system, assay, and
examination will be graded for level of complexity by assigning scores of 1, 2, or 3
within each criteria. The score of “1” indicates the lowest level of complexity, and
the score of “3” indicates the highest level. These scores will be totaled. Test
systems, assays or examinations receiving scores of 12 or less will be categorized as
moderate complexity, while those receiving scores above 12 will be categorized as
high complexity.
NOTE: A score of “2” will be assigned to a criteria heading when the
characteristics for a particular test are intermediate between the description listed
for scores of “1” and “3.”
(1) Knowledge.
(i) Score 1.
(A) Minimal scientific and technical knowledge is required to perform the
test; and
(B) Knowledge required to perform the test may be obtained through on-the-job instruction.
(ii) Score 3. Specialized scientific and technical knowledge is essential to
perform preanalytic, analytic or postanalytic phases of the testing.
(2) Training and experience.
(i) Score 1.
(A) Minimal training is required for preanalytic, analytic and postanalytic
phases of the testing process; and
(B) Limited experience is required to perform the test.
(ii) Score 3.
(A) Specialized training is essential to perform the preanalytic, analytic or
postanalytic testing process; or
(B) Substantial experience may be necessary for analytic test performance.
(3) Reagents and materials preparation.
(i) Score 1.
(A) Reagents and materials are generally stable and reliable; and
(B) Reagents and materials are prepackaged, or premeasured, or require no
special handling, precautions or storage conditions.
(ii) Score 3.
(A) Reagents and materials may be labile and may require special handling
to assure reliability; or
(B) Reagents and materials preparation may include manual steps such as
gravimetric or volumetric measurements.
(4) Characteristics of operational steps.
(i) Score l. Operational steps are either automatically executed (such as
pipetting, temperature monitoring, or timing of steps), or are easily
controlled.
(ii) Score 3. Operational steps in the testing process require close monitoring
or control, and may require special specimen preparation, precise
temperature control or timing of procedural steps, accurate pipetting, or
extensive calculations.
(5) Calibration, quality control, and proficiency testing materials.
(i) Score 1.
(A) Calibration materials are stable and readily available;
(B) Quality control materials are stable and readily available; and
(C) External proficiency testing materials, when available, are stable.
(ii) Score 3.
(A) Calibration materials, if available, may be labile;
(B) Quality control materials may be labile, or not available; or
(C) External proficiency testing materials, if available, may be labile.
(6) Test system troubleshooting and equipment maintenance.
(i) Score l.
(A) Test system troubleshooting is automatic or self-correcting, or clearly
described or requires minimal judgment; and
(B) Equipment maintenance is provided by the manufacturer, is seldom
needed, or can easily be performed.
(ii) Score 3.
(A) Troubleshooting is not automatic and requires decision-making and
direct intervention to resolve most problems; or
(B) Maintenance requires special knowledge, skills, and abilities.
(7) Interpretation and judgment.
(i) Score 1.
(A) Minimal interpretation and judgment are required to perform
preanalytic, analytic and postanalytic processes; and
(B) Resolution of problems requires limited independent interpretation and
judgment; and
(ii) Score 3.
(A) Extensive independent interpretation and judgment are required to
perform the preanalytic, analytic or postanalytic processes; and
(B) Resolution of problems requires extensive interpretation and judgment.
(b) Revisions to the criteria for categorization
The Clinical Laboratory Improvement Advisory Committee, as defined in subpart
T of this part, will conduct reviews upon request of HHS and recommend to HHS
revisions to the criteria for categorization of tests.
(c) Process for device/test categorization utilizing the scoring system
under §493.17(a). (1)(i) For new commercial test systems, assays, or
examinations, the manufacturer, as part of its 510(k) and PMA application to FDA,
will submit supporting data for device/test categorization. FDA will determine the
complexity category, notify the manufacturers directly, and will simultaneously
inform both CMS and CDC of the device/test category. FDA will consult with CDC
concerning test categorization in the following three situations:
(A) When categorizing previously uncategorized new technology;
(B) When FDA determines it to be necessary in cases involving a request for
a change in categorization; and
(C) If a manufacturer requests review of a categorization decision by FDA in
accordance with 21 CFR 10.75.
(ii) Test categorization will be effective as of the notification to the applicant.
(2) For test systems, assays, or examinations not commercially available, a
laboratory or professional group may submit a written request for categorization to
PHS. These requests will be forwarded to CDC for evaluation; CDC will determine
complexity category and notify the applicant, CMS, and FDA of the categorization
decision. In the case of request for a change of category or for previously
uncategorized new technology, PHS will receive the request application and forward
it to CDC for categorization.
(3) A request for recategorization will be accepted for review if it is based on new
information not previously submitted in a request for categorization or
recategorization by the same applicant and will not be considered more frequently
than once per year.
(4) If a laboratory test system, assay or examination does not appear on the lists of
tests in the FEDERAL REGISTER notices, it is considered to be a test of high
complexity until PHS, upon request, reviews the matter and notifies the applicant of
its decision. Test categorization is effective as of the notification to the applicant.
(5) PHS will publish revisions periodically to the list of moderate and high
complexity tests in the FEDERAL REGISTER in a notice with opportunity for
comment.
Interpretive Guidelines §493.17(c)(5)
To determine which tests are categorized as waived or nonwaived (i.e., moderate or high
complexity tests), refer to the following web link for the FDA categorization database
(http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfCLIA/Search.cfm?sAN=0). Test
systems, assays, and examinations not yet classified are considered high complexity.
Significant deficiencies cited under this condition may also indicate deficiencies under
personnel responsibilities.
NOTE: A modified waived or moderate complexity test (including modifications in its
intended use) is considered uncategorized for CLIA purposes and therefore becomes a
high complexity test.
§493.19 Provider-performed microscopy (PPM) procedures
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) Requirement. To be categorized as a PPM procedure, the procedure must meet
the criteria specified in paragraph (b) of this section.
(b) Criteria. Procedures must meet the following specifications:
(1) The examination must be personally performed by one of the following
practitioners:
(i) A physician during the patient’s visit on a specimen obtained from his or her own
patient or from a patient of a group medical practice of which the physician is a
member or an employee.
(ii) A midlevel practitioner, under the supervision of a physician or in independent
practice only if authorized by the State, during the patient’s visit on a specimen
obtained from his or her own patient or from a patient of a clinic, group medical
practice, or other health care provider of which the midlevel practitioner is a
member or an employee.
(iii) A dentist during the patient’s visit on a specimen obtained from his or her own
patient or from a patient of a group dental practice of which the dentist is a member
or an employee.
(2) The procedure must be categorized as moderately complex.
(3) The primary instrument for performing the test is the microscope, limited to
bright-field or phase-contrast microscopy.
(4) The specimen is labile or delay in performing the test could compromise the
accuracy of the test result.
(5) Control materials are not available to monitor the entire testing process.
Interpretive Guidelines for 493.19(b)(5)
Refer to the interpretive guidelines at 493.1256(h) for more guidance on PPM quality
control.
(6) Limited specimen handling or processing is required.
(c) Provider-performed microscopy (PPM) examinations. A laboratory may qualify
to perform tests under this section if it restricts PPM examinations to one or more of
the following procedures (or additional procedures added to this list as provided
under paragraph (d) of this section), waived tests and no others:
(1) All direct wet mount preparations for the presence or absence of bacteria,
fungi, parasites, and human cellular elements.
Interpretive Guideline for 493.19(c)(1)
All direct wet mounts for scabies (which includes mineral oil preparations) are PPM
examinations. All direct wet mount preparations for the presence or absence of bacteria,
fungi, parasites, and human cellular elements include the direct wet mount examinations
for Clue cells. Clue cells are vaginal epithelial cells covered with bacteria.
(2) All potassium hydroxide (KOH) preparations.
Interpretive Guideline for 493.19(c)(2)
All KOH preparations include KOH that is mixed directly with a stain or the stain is a
drop added directly to the KOH preparation and the slide is examined with bright-field
microscopy.
(3) Pinworm examinations.
(4) Fern tests.
(5) Post-coital direct, qualitative examinations of vaginal or cervical mucous.
(6) Urine sediment examinations.
Interpretive Guideline for 493.19(c)(6)
It is acceptable in a PPM laboratory to add stain to resuspend the urine sediment to assist
in identifying the formed elements in the urine.
(7) Nasal smears for granulocytes.
(8) Fecal leukocyte examinations.
(9) Qualitative semen analysis (limited to the presence or absence of sperm and
detection of motility).
(d) Revision to criteria and the list of PPM procedures
(1) The CLIAC conducts reviews upon HHS’ request and recommends to HHS
revisions to the criteria for categorization of procedures.
(2) HHS determines whether a laboratory procedure meets the criteria listed under
paragraph (b) of this section for a PPM procedure. Revisions to the list of PPM
procedures proposed by HHS are published in the FEDERAL REGISTER as a
notice with an opportunity for public comment.
(e) Laboratory requirements
Laboratories eligible to perform PPM examinations must--
(1) Meet the applicable requirements in subpart C or subpart D, and subparts F, H,
J, K, and M of this part.
(2) Be subject to inspection as specified under subpart Q of this part.
§493.20 Laboratories performing tests of moderate complexity
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) A laboratory may qualify for a certificate to perform tests of moderate
complexity provided that it restricts its test performance to waived tests or
examinations and one or more tests or examinations meeting criteria for tests
of moderate complexity including the subcategory of PPM procedures.
(b) A laboratory that performs tests or examinations of moderate complexity
must meet the applicable requirements in subpart C or subpart D, and
subparts F, H, J, K, M, and Q of this part. Under a registration certificate or
certificate of compliance, laboratories also performing PPM procedures must
meet the inspection requirements at §§493.1773 and 493.1777.
(c) If the laboratory also performs waived tests, compliance with § 493.801(a)
and (b)(7) and subparts J, K, and M of this part is not applicable to the
waived tests. However, the laboratory must comply with the requirements in
§§ 493.15(e), 493.801(b)(1) through (6), 493.1771, 493.1773, and 493.1775.
§493.25 Laboratories performing tests of high complexity
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) A laboratory must obtain a certificate for tests of high complexity if it
performs one or more tests that meet the criteria for tests of high complexity
as specified in §493.17(a).
(b) A laboratory performing one or more tests of high complexity must meet the
applicable requirements of subpart C or subpart D, and subparts F, H, J, K,
M, and Q of this part.
(c) If the laboratory also performs tests of moderate complexity, the applicable
requirements of subparts H, J, K, M, and Q of this part must be met. Under
a registration certificate or certificate of compliance, PPM procedures must
meet the inspection requirements in §§493.1773 and 493.1777.
(d) If the laboratory also performs waived tests, compliance with §§ 493.801(a)
and 493.801(b)(7) and subparts J, K, and M of this part are not applicable to
the waived tests. However, the laboratory must comply with the
requirements in §§493.15(e), 493.801(b)(1) through (6), 493.1771, 493.1773
and 493.1775.
Subpart B--Certificate of Waiver
§493.35 Application for a certificate of waiver
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) Filing of application.
Except as specified in paragraph (b) of this section, a laboratory performing only
one or more waived tests listed in §493.15 must file a separate application for each
laboratory location.
Interpretive Guidelines §493.35(a)
NOTE: See section 6036 of the SOM for instructions on handling a laboratory operating
without a CLIA certificate. Per policy letter QSO18-20-CLIA, multiple laboratories with
separate CLIA numbers may operate at one location as long as it can be demonstrated
that each laboratory is operating as a separate and distinct entity.
(b) Exceptions
(1) Laboratories that are not at a fixed location, that is, laboratories that move
from testing site to testing site, such as mobile units providing laboratory testing,
health screening fairs, or other temporary testing locations may be covered
under the certificate of the designated primary site or home base, using its
address.
Interpretive Guidelines §493.35(b)(1)
A mobile unit is a laboratory located within a self-contained vehicle, such as a van. The
vehicle moves from location to location to perform laboratory testing activities. Mobile
vans will be distinguished by the vehicle identification number (VIN#).
If a mobile laboratory operates in more than one State and does not obtain a separate
certificate from each State, contact CMS to determine which State conducts the
inspection. See section 6010.1.1 of the SOM for additional information on mobile
laboratories.
Each laboratory that moves from testing site to testing site, or has a temporary testing
location, should provide the survey agency with the home base or central dispatch phone
number and the locations where additional testing is performed.
A temporary testing site is considered a location not used to permanently house
instruments, equipment, personnel and records, e.g., a health fair. See section 6010.1.2 of
the SOM for further guidance.
See section 6010.1.2.1 of the SOM for guidance for Home Health Agencies with multiple
sites.
(2) Not-for-profit or Federal, State, or local government laboratories that
engage in limited (not more than a combination of 15 moderately complex or
waived tests per certificate) public health testing may file a single application.
Interpretive Guidelines §493.35(b)(2)
See section 6010.2 of the SOM for the definition for limited public health testing. Note
that laboratories operating under a certificate of waiver may not perform moderate or
high complexity testing.
(3) Laboratories within a hospital that are located at contiguous buildings on the
same campus and under common direction may file a single application or
multiple applications for the laboratory sites within the same physical
location or street address.
Interpretive Guidelines §493.35(b)(3)
Common direction means that all testing sites are under one designated director.
Street address is the address assigned by the Post Office and is the physical location of
the laboratory. The street address may be different from the mailing address, which can
be a Post Office box or a billing address. For large hospitals, such as a university campus
facility, that may contain laboratories in separate buildings, consult with CMS to
determine if the hospital is eligible for a single certificate.
(c) Application format and contents
The application must--
(1) Be made to HHS or its designee on a form or forms prescribed by HHS;
(2) Be signed by an owner, or by an authorized representative of the
laboratory who attests that the laboratory will be operated in accordance
with requirements established by the Secretary under section 353 of the
PHS Act; and
(3) Describe the characteristics of the laboratory operation and the
examinations and other test procedures performed by the laboratory
including--
(i) The name and the total number of test procedures and examinations
performed annually (excluding tests the laboratory may run for quality
control, quality assurance or proficiency testing purposes;
(ii) The methodologies for each laboratory test procedure or examination
performed, or both; and
(iii) The qualifications (educational background, training, and experience) of
the personnel directing and supervising the laboratory and performing
the laboratory examinations and test procedures.
(d) Access requirements
Laboratories that perform one or more waived tests listed in §493.15(c) and no
other tests must meet the following conditions:
Interpretive Guidelines §493.35(d)
Cite deficiencies for not following manufacturer’s instructions at §493.15(e). (Use
D1001)
(1) Make records available and submit reports to HHS as HHS may reasonably
require to determine compliance with this section and §493.15(e);
(2) Agree to permit announced and unannounced inspections by HHS in
accordance with subpart Q of this part under the following circumstances:
(i) When HHS has substantive reason to believe that the laboratory is being
operated in a manner that constitutes an imminent and serious risk to
human health.
Interpretive Guidelines §493.35(d)(2)(i)
Consult with CMS for assistance in determining when there is substantive reason to
believe that the laboratory is being operated in a manner that constitutes an imminent and
serious risk to human health.
An example of a substantive reason to inspect waived testing is if testing personnel are
observed cutting urine dipsticks in half. (This violates both the manufacturer’s
instructions and causes questionable results to be reported.)
(ii) To evaluate complaints from the public.
(iii) On a random basis to determine whether the laboratory is performing
tests not listed in §493.15.
Interpretive Guidelines §493.35(d)(2)(ii)-(iii)
NOTE: See Chapter 5 of the SOM for specific procedures regarding complaint
investigations.
(iv) To collect information regarding the appropriateness of waiver of tests
listed in §493.15.
(e) Denial of application
If HHS determines that the application for a certificate of waiver is to be denied,
HHS will--
(1) Provide the laboratory with a written statement of the grounds on which
the denial is based and an opportunity for appeal, in accordance with the
procedures set forth in subpart R of this part;
(2) Notify a laboratory that has its application for a certificate of waiver
denied that it cannot operate as a laboratory under the PHS Act unless
the denial is overturned at the conclusion of the administrative appeals
process provided by subpart R; and
(3) Notify the laboratory that it is not eligible for payment under the
Medicare and Medicaid programs.
§493.37 Requirements for a certificate of waiver
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) HHS will issue a certificate of waiver to a laboratory only if the laboratory meets
the requirements of §493.35.
(b) Laboratories issued a certificate of waiver--
(1) Are subject to the requirements of this subpart and §493.15(e) of subpart A of
this part; and
Interpretive Guidelines §493.37(b)(1)
Cite the laboratory’s failure to follow manufacturer’s instructions at §493.15(e). (Use
D1001.)
(2) Must permit announced or unannounced inspections by HHS in accordance with
subpart Q of this part.
(c) Laboratories must remit the certificate of waiver fee specified in subpart F of
this part.
(d) In accordance with subpart R of this part, HHS will suspend or revoke or limit a
laboratory’s certificate of waiver for failure to comply with the requirements of this
subpart. In addition, failure to meet the requirements of this subpart will result in
suspension or denial of payments under Medicare and Medicaid in accordance with
subpart R of this part.
Interpretive Guidelines §493.37(d)
NOTE: See the Adverse Action section of the SOM beginning at section 6250 for
enforcement procedures.
(e)(1) A certificate of waiver issued under this subpart is valid for no more than 2
years. In the event of a non-compliance determination resulting in HHS action to
revoke, suspend, or limit the laboratory’s certificate of waiver, HHS will provide the
laboratory with a statement of grounds on which the determination of noncompliance is based and offer an opportunity for appeal as provided in subpart R of
this part.
(2) If the laboratory requests a hearing within the time specified by HHS, it retains
its certificate of waiver or reissued certificate of waiver until a decision is made by
an administrative law judge, as specified in subpart R of this part, except when
HHS finds that conditions at the laboratory pose an imminent and serious risk to
human health.
(3) For laboratories receiving payment from the Medicare or Medicaid program,
such payments will be suspended on the effective date specified in the notice to the
laboratory of a non-compliance determination even if there has been no appeals
decision issued.
(f) A laboratory seeking to renew its certificate of waiver must--
(1) Complete the renewal application prescribed by HHS and return it to HHS not
less than 9 months nor more than 1 year before the expiration of the certificate; and
(2) Meet the requirements of §§493.35 and 493.37.
§493.37(g) A laboratory with a certificate of waiver that wishes to perform
examinations or tests not listed in the waiver test category must meet the
requirements set forth in subpart C or subpart D of this part, as applicable.
§493.39 Notification requirements for laboratories issued a certificate of
waiver
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
Laboratories performing one or more tests listed in §493.15 and no others must
notify HHS or its designee--
(a) Before performing and reporting results for any test or examination that is not
specified under §493.15 for which the laboratory does not have the appropriate
certificate as required in subpart C or subpart D of this part, as applicable; and
Interpretive Guidelines §493.39(a)
NOTE: See section 6036 of the SOM for instructions on handling a laboratory operating
without an appropriate CLIA certificate.
(b) Within 30 days of any change(s) in--
(1) Ownership;
(2) Name;
(3) Location; or
(4) Director.
Interpretive Guidelines §493.39(b)
NOTE: See section 6016 of the SOM for applicable instructions on handling changes in
ownership, name, location, or director.
Because there is no D-tag assigned to this regulation, you may use D0000 to cite noncompliance with this requirement.
Subpart C--Registration Certificate, Certificate for Provider-Performed Microscopy Procedures, and Certificate of
Compliance
§493.43 Application for registration certificate, certificate for provider-performed microscopy (PPM) procedures, and certificate of compliance
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) Filing of application
Except as specified in paragraph (b) of this section, all laboratories performing
nonwaived testing must file a separate application for each laboratory location.
Interpretive Guidelines §493.43(a)
NOTE: See section 6036 of the SOM for instructions on handling a laboratory operating
without a CLIA certificate. Per policy letter QSO 18-20 CLIA, multiple laboratories
with separate CLIA numbers may operate at one location as long as it can be
demonstrated that each laboratory is operating as a separate and distinct entity.
(b) Exceptions
(1) Laboratories that are not at a fixed location, that is, laboratories that move from
testing site to testing site, such as mobile units providing laboratory testing, health
screening fairs, or other temporary testing locations may be covered under the
certificate of the designated primary site or home base, using its address.
Interpretive Guidelines §493.43(b)(1)
A mobile unit is a laboratory located within a self-contained vehicle, such as a van. The
vehicle moves from location to location to perform laboratory testing activities. Mobile
vans will be distinguished by the vehicle identification number (VIN#).
If a mobile laboratory operates in more than one State and does not obtain a separate
certificate for each State, contact CMS to determine which State conducts the inspection.
See section 6010.1.1 of the SOM for additional information on mobile laboratories.
Each laboratory that moves from testing site to testing site, or has a temporary testing
location, should provide the survey agency with the home base or central dispatch phone
number and the locations where additional testing is performed.
A temporary testing site is considered a location not used to permanently house
instruments, equipment, personnel and records, e.g., a health fair. See §6036.3 of the
SOM for further guidance.
NOTE: See section 6010.1.2.1 of the SOM for guidance for home health agencies with
multiple sites.
(2) Not-for-profit or Federal, State, or local government laboratories that engage in
limited (not more than a combination of 15 moderately complex or waived tests per
certificate) public health testing may file a single application.
Interpretive Guidelines §493.43(b)(2)
NOTE: See section 6010.2 of the SOM for information on limited public health testing.
(3) Laboratories within a hospital that are located at contiguous buildings on the
same campus and under common direction may file a single application or multiple
applications for the laboratory sites within the same physical location or street
address.
Interpretive Guidelines §493.43(b)(3)
In instances where the main laboratory is certified to perform waived, moderate and/or
high complexity tests, the alternate sites may perform testing in all complexities covered
by the certificate provided that all other applicable requirements are met (e.g., quality
control, personnel).
Common direction means that all sites are under one designated director.
Street address is the address assigned by the Post Office and is the physical location of
the laboratory. The street address may be different from the mailing address, which can
be a Post Office box or a billing address. For large hospitals, such as a university campus
facility, that may contain laboratories in separate buildings, consult with CMS to
determine if the hospital is eligible for a single certificate.
(c) Application format and contents
The application must--(1) Be made to HHS or its designee on a form or forms
prescribed by HHS;
(2) Be signed by an owner, or by an authorized representative of the laboratory who
attests that the laboratory will be operated in accordance with the requirements
established by the Secretary under section 353 of the Public Health Service Act; and
(3) Describe the characteristics of the laboratory operation and the examinations
and other test procedures performed by the laboratory including--
(i) The name and total number of test procedures and examinations
performed annually (excluding waived tests or tests for quality control,
quality assurance or proficiency testing purposes);
(ii) The methodologies for each laboratory test procedure or examination
performed, or both;
(iii) The qualifications (educational background, training, and experience) of
the personnel directing and supervising the laboratory and performing
the examinations and test procedures.
(d) Access and reporting requirements
All laboratories must make records available and submit reports to HHS as HHS
may reasonably require to determine compliance with this section.
§493.45 Requirements for a registration certificate
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
Laboratories performing only waived tests, PPM procedures, or any combination of
these tests, are not required to obtain a registration certificate.
(a) A registration certificate is required—
(1) Initially for all laboratories performing test procedures of moderate complexity
(other than the subcategory of PPM procedures) or high complexity, or both; and
(2) For all laboratories that have been issued a certificate of waiver or certificate
for PPM procedures that intend to perform tests of moderate or high complexity, or
both, in addition to those tests listed in §493.15 (c) or specified as PPM procedures.
Interpretive Guidelines §493.45(a)
All facilities performing laboratory testing must have a registration, compliance or
accreditation certificate, a certificate for provider-performed microscopy procedures or a
certificate of waiver prior to performing patient testing.
NOTE: See section 6036 of the SOM for instructions on handling a laboratory operating
without an appropriate CLIA certificate.
(b) HHS will issue a registration certificate if the laboratory--
(1) Complies with the requirements of §493.43;
(2) Agrees to notify HHS or its designee within 30 days of any changes in ownership,
name, location, director or technical supervisor (laboratories performing high
complexity testing only);
(3) Agrees to treat proficiency testing samples in the same manner as it treats
patient specimens; and
(4) Remits the fee for the registration certificate, as specified in subpart F of this
part.
(c) Prior to the expiration of the registration certificate, a laboratory must--
(1) Remit the certificate fee specified in subpart F of this part;
(2) Be inspected by HHS as specified in subpart Q of this part; and
(3) Demonstrate compliance with the applicable requirements of this subpart and
subparts H, J, K, M, and Q of this part.
(d) In accordance with subpart R of this part, HHS will initiate suspension or
revocation of a laboratory’s registration certificate and will deny the laboratory’s
application for a certificate of compliance for failure to comply with the
requirements set forth in this subpart. HHS may also impose certain alternative
sanctions. In addition, failure to meet the requirements of this subpart will result in
suspension of payments under Medicare and Medicaid as specified in subpart R of
this part.
(e) A registration certificate is--
(1) Valid for a period of no more than two years or until such time as an inspection
to determine program compliance can be conducted, whichever is shorter; and
(2) Not renewable; however, the registration certificate may be reissued if
compliance has not been determined by HHS prior to the expiration date of the
registration certificate.
(f) In the event of a noncompliance determination resulting in an HHS denial of a
laboratory’s certificate of compliance application, HHS will provide the laboratory
with a statement of grounds on which the noncompliance determination is based
and offer an opportunity for appeal as provided in subpart R.
Interpretive Guidelines §493.45(f)
NOTE: See the Appeals section of the SOM beginning at sections 6258 and 6300 for
instructions on denial of a certificate application.
(g) If the laboratory requests a hearing within the time specified by HHS, it retains
its registration certificate or reissued registration certificate until a decision is made
by an administrative law judge as provided in subpart R of this part, except when
HHS finds that conditions at the laboratory pose an imminent and serious risk to
human health.
(h) For laboratories receiving payment from the Medicare or Medicaid program,
such payments will be suspended on the effective date specified in the notice to the
laboratory of denial of the certificate application even if there has been no appeals
decision issued.
§493.47 Requirements for a certificate for provider-performed
microscopy (PPM) procedures.
(Rev. 236; Issued: 01-23-26; Effective: 09-12-25; Implementation: 09-12-25)
(a) A certificate for PPM procedures is required-
(1) Initially for all laboratories performing test procedures specified as PPM
procedures; and
(2) For all certificate of waiver laboratories that intend to perform only test
procedures specified as PPM procedures in addition to those tests listed in
§493.15(c).
(b) HHS will issue a certificate for PPM procedures if the laboratory-
(1) Complies with the requirements of §493.43; and
(2) Remits the fee for the certificate, as specified in subpart F of this part.
(c) Laboratories issued a certificate for PPM procedures are subject to-
(1) The notification requirements of §493.53;
(2) The applicable requirements of this subpart and subparts H, J, K, and M of this
part; and
(3) Inspection only under the circumstances specified under §§493.1773 and
493.1775, but are not routinely inspected to determine compliance with the
requirements specified in paragraphs (c)(1) and (2) of this section.
(d) In accordance with subpart R of this part, HHS will initiate suspension,
limitation, or revocation of a laboratory’s certificate for PPM procedures for failure
to comply with the applicable requirements set forth in this subpart. HHS may also
impose certain alternative sanctions. In addition, failure to meet the requirements of
this subpart may result in suspension of all or part of payments under Medicare and
Medicaid, as specified in subpart R of this part.
(e) A certificate for PPM procedures is valid for a period of no more than 2 years.
§493.49 Requirements for a certificate of compliance.
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
A certificate of compliance may include any combination of tests categorized as high
complexity or moderate complexity or listed in §493.15(c) as waived tests. Moderate
complexity tests may include those specified as PPM procedures.
(a) HHS will issue a certificate of compliance to a laboratory only if the laboratory--
(1) Meets the requirements of §§493.43 and 493.45;
(2) Remits the certificate fee specified in subpart F of this part; and
(3) Meets the applicable requirements of this subpart and subparts H, J, K, M, and
Q of this part.
(b) Laboratories issued a certificate of compliance--
(1) Are subject to the notification requirements of §493.51; and
(2) Must permit announced or unannounced inspections by HHS in accordance with
subpart Q of this part--
(i) To determine compliance with the applicable requirements of this part;
(ii) To evaluate complaints;
(iii) When HHS has substantive reason to believe that tests are being performed, or
the laboratory is being operated in a manner that constitutes an imminent and
serious risk to human health; and
(iv) To collect information regarding the appropriateness of tests listed in §493.15 or
tests categorized as moderate complexity (including the subcategory) or high
complexity.
(c) Failure to comply with the requirements of this subpart will result in--
(1) Suspension, revocation or limitation of a laboratory’s certificate of compliance in
accordance with subpart R of this part; and
(2) Suspension or denial of payments under Medicare and Medicaid in accordance
with subpart R of this part.
(d) A certificate of compliance issued under this subpart is valid for no more than 2
years.
(e) In the event of a noncompliance determination resulting in an HHS action to
revoke, suspend or limit the laboratory’s certificate of compliance, HHS will--
(1) Provide the laboratory with a statement of grounds on which the determination
of noncompliance is based; and
(2) Offer an opportunity for appeal as provided in subpart R of this part. If the
laboratory requests a hearing within 60 days of the notice of sanction, it retains its
certificate of compliance or reissued certificate of compliance until a decision is
made by an administrative law judge (ALJ) as provided in subpart R of this part,
except when HHS finds that conditions at the laboratory pose an imminent and
serious risk to human health or when the criteria at §493.1840(a)(4) and (5) are met.
(f) For laboratories receiving payment from the Medicare or Medicaid program,
such payments will be suspended on the effective date specified in the notice to the
laboratory of a noncompliance determination even if there has been no appeals
decision issued.
(g) A laboratory seeking to renew its certificate of compliance must--
(1) Complete and return the renewal application to HHS 9 to 12 months prior to the
expiration of the certificate of compliance; and
(2) Meet the requirements of §493.43 and paragraphs (a)(2) and (b)(2) of this
section.
(h) If HHS determines that the application for the renewal of a certificate of
compliance must be denied or limited, HHS will notify the laboratory in writing of
the--
(1) Basis for denial of the application; and
(2) Opportunity for appeal as provided in subpart R of this part.
Interpretive Guidelines §493.49(h)(2)
NOTE: See the Appeals section of the SOM beginning at sections 6258 and 6300 for
instructions on denial of a certificate application.
(i) If the laboratory requests a hearing within the time period specified by HHS, the
laboratory retains its certificate of compliance or reissued certificate of compliance
until a decision is made by an ALJ as provided in subpart R, except when HHS
finds that conditions at the laboratory pose an imminent and serious risk to human
health.
(j) For laboratories receiving payment from the Medicare or Medicaid program,
such payments will be suspended on the effective date specified in the notice to the
laboratory of nonrenewal of the certificate of compliance even if there has been no
appeals decision issued.
§493.51 Notification requirements for laboratories issued a certificate of
compliance
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
Laboratories issued a certificate of compliance must meet the following conditions:
(a) Notify HHS or its designee within 30 days of any change in--
(1) Ownership;
(2) Name;
(3) Location;
(4) Director; or
(5) Technical supervisor (laboratories performing high complexity only).
(b) Notify HHS no later than 6 months after performing any test or examination
within a specialty or subspecialty area that is not included on the laboratory’s
certificate of compliance, so that compliance with requirements can be determined.
(c) Notify HHS no later than 6 months after any deletions or changes in test
methodologies for any test or examination included in a specialty or subspecialty, or
both, for which the laboratory has been issued a certificate of compliance.
Interpretive Guidelines §493.51(a)-(c)
See the section of the SOM beginning at section 6016 for handling changes in ownership,
name, location, personnel and test methodology, or additions or deletions of specialties or
subspecialties that may result in changes in complexity levels for the laboratory.
See the SOM beginning at section 6260 for instructions on handling laboratories that are
going out of business or voluntarily withdrawing from all testing, including laboratories
that are taking these actions while an enforcement action is pending.
Because there is no D-tag assigned to this regulation, you may use D0000 to cite noncompliance with this requirement.
§493.53 Notification requirements for laboratories issued a certificate
for provider-performed microscopy (PPM) procedures
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
Laboratories issued a certificate for PPM procedures must notify HHS or its
designee--
(a) Before performing and reporting results for any test of moderate or high
complexity, or both, in addition to tests specified as PPM procedures or any test or
examination that is not specified under §493.15(c), for which it does not have a
registration certificate as required in subpart C or subpart D, as applicable, of this
part; and
(b) Within 30 days of any change in--
(1) Ownership;
(2) Name;
(3) Location; or
(4) Director
Interpretive Guidelines §493.53(b)
See the section of the SOM beginning at section 6016 for handling changes in ownership,
name, location, personnel and test methodology, or additions or deletions of specialties or
subspecialties that may result in changes in complexity levels for the laboratory.
See the SOM beginning at section 6260 for instructions on handling laboratories that are
going out of business or voluntarily withdrawing from all testing, including laboratories
that are taking these actions while an enforcement action is pending.
Because there is no D-tag assigned to this regulation, you may use D0000 to cite noncompliance with this requirement.
Subpart D--Certificate of Accreditation
§493.55 Application for registration certificate and certificate of
accreditation
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) Filing of application
A laboratory may be issued a certificate of accreditation in lieu of the applicable
certificate specified in subpart B or subpart C of this part provided the laboratory--
(1) Meets the standards of a private non-profit accreditation program
approved by HHS in accordance with subpart E; and
Interpretive Guidelines §493.55(a)(1)
When HHS approves accreditation organizations and State licensure programs, CMS is
notified and the approved organizations and programs are published as a notice in the
FEDERAL REGISTER.
NOTE: See sections 6150-6151 of the SOM.
(2) Files a separate application for each location, except as specified in
paragraph (b) of this section.
Interpretive Guidelines §493.55(a)(2)
See section 6036 of the SOM for instructions on handling a laboratory operating without
a CLIA certificate. Per policy letter QSO 18-20 CLIA, multiple laboratories with
separate CLIA numbers may operate at one location as long as it can be demonstrated
that each laboratory is operating as a separate and distinct entity.
(b) Exceptions
(1) Laboratories that are not at fixed locations, that is, laboratories that move
from testing site to testing site, such as mobile units providing laboratory
testing, health screening fairs, or other temporary testing locations may
be covered under the certificate of the designated primary site or home
base, using its address.
Interpretive Guidelines §493.55(b)(1)
A mobile unit is a laboratory located within a self-contained vehicle, such as a van. The
vehicle moves from location to location to perform laboratory testing activities. Mobile
vans will be distinguished by the vehicle identification number (VIN#).
If a mobile laboratory operates in more than one State and does not obtain a separate
certificate from each State, contact CMS to determine which State conducts the
inspection. See section 6010.1.1 of the SOM for additional information on mobile
laboratories.
Each laboratory that moves from testing site to testing site, or has a temporary testing
location, should provide the survey agency with the home base or central dispatch phone
number and the locations where additional testing is performed.
A temporary testing site is considered a location not used to permanently house
instruments, equipment, personnel and records, e.g., a health fair. See section 6036.3 of
the SOM for further guidance.
See section 6010.1.2.1 of the SOM for guidance for home health agencies with multiple
sites.
(2) Not-for-profit or Federal, State, or local government laboratories that
engage in limited (not more than a combination of 15 moderately complex
or waived tests per certificate) public health testing may file a single
application.
Interpretive Guidelines §493.55(b)(2)
See section 6010.2 of the SOM for the definition of limited public health testing.
(3) Laboratories within a hospital that are located at contiguous buildings on
the same campus and under common direction may file a single
application or multiple applications for the laboratory sites within the
same physical location or street address.
Interpretive Guidelines §493.55(b)(3)
Common direction means that all sites are under one designated director.
Street address is the address assigned by the Post Office and is the physical location of
the laboratory. The street address may be different from the mailing address, which can
be a Post Office box or a billing address. For large hospitals, such as a university campus
facility, that may contain laboratories in separate buildings, consult with CMS to
determine if the hospital is eligible for a single certificate.
(c) Application format and contents
The application must--
(1) Be made to HHS on a form or forms prescribed by HHS;
(2) Be signed by an owner or authorized representative of the laboratory who
attests that the laboratory will be operated in accordance with the
requirements established by the Secretary under section 353 of the Public
Health Service Act; and
(3) Describe the characteristics of the laboratory operation and the
examinations and other test procedures performed by the laboratory
including--
(i) The name and total number of tests and examinations performed
annually (excluding waived tests and tests for quality control, quality
assurance or proficiency testing purposes);
(ii)The methodologies for each laboratory test procedure or examination
performed, or both; and
(iii)The qualifications (educational background, training, and experience)
of the personnel directing and supervising the laboratory and
performing the laboratory examinations and test procedures.
(d) Access and reporting requirements.
All laboratories must make records available and submit reports to HHS as HHS
may reasonably require to determine compliance with this section.
§493.57 Requirements for a registration certificate
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
A registration certificate is required for all laboratories seeking a certificate of
accreditation, unless the laboratory holds a valid certificate of compliance issued by
HHS.
Interpretive Guidelines §493.57
See section 6036 of the SOM for instructions on handling a laboratory operating without
a CLIA certificate.
§493.57(a) HHS will issue a registration certificate if the laboratory--
(1) Complies with the requirements of §493.55;
(2) Agrees to notify HHS within 30 days of any changes in ownership, name,
location, director, or supervisor (laboratories performing high complexity testing
only);
(3) Agrees to treat proficiency testing samples in the same manner as it treats
patient specimens; and
(4) Remits the fee for the registration certificate specified in subpart F of this part.
(b)(1) The laboratory must provide HHS with proof of accreditation by an approved
accreditation program--
(i) Within 11 months of issuance of the registration certificate; or
(ii) Prior to the expiration of the certificate of compliance.
(2) If such proof of accreditation is not supplied within this timeframe, the
laboratory must meet, or continue to meet, the requirements of §493.49.
(c) In accordance with subpart R of this part, HHS will initiate suspension,
revocation, or limitation of a laboratory’s registration certificate and will deny the
laboratory’s application for a certificate of accreditation for failure to comply with
the requirements set forth in this subpart. In addition, failure to meet the
requirements of this subpart will result in suspension or denial of payments under
Medicare and Medicaid as specified in subpart R of this part.
(d) A registration certificate is valid for a period of no more than 2 years. However,
it may be reissued if the laboratory is subject to subpart C of this part, as specified
in §493.57(b)(2) and compliance has not been determined by HHS before the
expiration date of the registration certificate.
(e) In the event that the laboratory does not meet the requirements of this subpart,
HHS will--
Interpretive Guidelines §493.57(c)-(e)
See the Appeals section of the SOM beginning at sections 6258 and 6300 for instructions
on denial of a certificate of accreditation application.
(1) Deny a laboratory’s request for certificate of accreditation;
(2) Notify the laboratory if it must meet the requirements for a certificate as defined
in subpart C of this part;
(3) Provide the laboratory with a statement of grounds on which the application
denial is based;
(4) Offer an opportunity for appeal on the application denial as provided in subpart
R of this part. If the laboratory requests a hearing within the time specified by HHS,
the laboratory will retain its registration certificate or reissued registration
certificate until a decision is made by an administrative law judge as provided in
subpart R, unless HHS finds that conditions at the laboratory pose an imminent and
serious risk to human health; and
(5) For those laboratories receiving payment from the Medicare or Medicaid
program, such payments will be suspended on the effective date specified in the
notice to the laboratory of denial of the request even if there has been no appeals
decision issued.
§493.61 Requirements for a certificate of accreditation
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
(a) HHS will issue a certificate of accreditation to a laboratory if the laboratory--
(1) Meets the requirements of §493.57 or, if applicable, §493.49 of subpart C
of this part; and
(2) Remits the certificate of accreditation fee specified in subpart F of this
part.
(b) Laboratories issued a certificate of accreditation must--
(1) Treat proficiency testing samples in the same manner as patient samples;
(2) Meet the requirements of §493.63;
(3) Comply with the requirements of the approved accreditation program;
(4) Permit random sample validation and complaint inspections as required
in subpart Q of this part;
(5) Permit HHS to monitor the correction of any deficiencies found through
the inspections specified in paragraph (b)(4) of this section;
Interpretive Guidelines §493.61(b)(5)
See sections 6166 and 6184 of the SOM for procedures on follow-up of correction of
deficiencies cited during validation inspections.
(6) Authorize the accreditation program to release to HHS the laboratory’s
inspection findings whenever HHS conducts random sample or complaint
inspections; and
(7) Authorize its accreditation program to submit to HHS the results of the
laboratory’s proficiency testing.
(c) A laboratory failing to meet the requirements of this section--
(1) Will no longer meet the requirements of this part by virtue of its
accreditation in an approved accreditation program;
(2) Will be subject to full determination of compliance by HHS;
(3) May be subject to suspension, revocation or limitation of the laboratory’s
certificate of accreditation or certain alternative sanctions; and
(4) May be subject to suspension of payments under Medicare and Medicaid
as specified in subpart R.
(d) A certificate of accreditation issued under this subpart is valid for no more than
2 years. In the event of a non-compliance determination as a result of a random
sample validation or complaint inspection, a laboratory will be subject to a full
review by HHS in accordance with §488.11 of this chapter.
Interpretive Guidelines §493.61(d)
42 CFR §488.11 lists State survey agency functions.
(e) Failure to meet the applicable requirements of part 493, will result in an action
by HHS to suspend, revoke or limit the certificate of accreditation. HHS will--
(1) Provide the laboratory with a statement of grounds on which the
determination of noncompliance is based;
(2) Notify the laboratory if it is eligible to apply for a certificate as defined in
subpart C of this part; and
(3) Offer an opportunity for appeal as provided in subpart R of this part.
(f) If the laboratory requests a hearing within the time frame specified by HHS--
(1) It retains its certificate of accreditation or reissued certificate of
accreditation until a decision is made by an administrative law judge as
provided in subpart R of this part, unless HHS finds that conditions at the
laboratory pose an imminent and serious risk to human health; and
(2) For those laboratories receiving payments from the Medicare or
Medicaid program, such payments will be suspended on the effective date
specified in the notice to the laboratory even if there has been no appeals
decision issued.
(g) In the event the accreditation organization’s approval is removed by HHS, the
laboratory will be subject to the applicable requirements of subpart C of this part or
§493.57.
Interpretive Guidelines §493.61(g)
Accrediting organizations which lose deemed status are required to notify their
participating laboratories. These laboratories must re-apply for accreditation with
another CMS-approved accrediting organization or apply for the appropriate CLIA
certificate with CMS.
(h) A laboratory seeking to renew its certificate of accreditation must--
(1) Complete and return the renewal application to HHS 9 to 12 months
prior to the expiration of the certificate of accreditation;
(2) Meet the requirements of this subpart; and
(3) Submit the certificate of accreditation fee specified in subpart F of this
part.
(i) If HHS determines that the renewal application for a certificate of
accreditation is to be denied or limited, HHS will notify the laboratory
in writing of--
(1) The basis for denial of the application;
(2) Whether the laboratory is eligible for a certificate as defined in
subpart C of this part;
(3) The opportunity for appeal on HHS’s action to deny the renewal
application for certificate of accreditation as provided in subpart R of
this part. If the laboratory requests a hearing within the time frame
specified by HHS, it retains its certificate of accreditation or reissued
certificate of accreditation until a decision is made by an
administrative law judge as provided in subpart R of this part, unless
HHS finds that conditions at the laboratory pose an imminent and
serious risk to human health; and
(4) Suspension of payments under Medicare or Medicaid for those laboratories
receiving payments under the Medicare or Medicaid programs.
§493.63 Notification requirements for laboratories issued a certificate of
accreditation
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
Laboratories issued a certificate of accreditation must:
(a) Notify HHS and the approved accreditation program within 30 days of any
changes in--
(1) Ownership;
(2) Name;
(3) Location; or
(4) Director.
(b) Notify the approved accreditation program no later than 6 months after
performing any test or examination within a specialty or subspecialty area that is
not included in the laboratory’s accreditation, so that the accreditation organization
can determine compliance and a new certificate of accreditation can be issued.
(c) Notify the accreditation program no later than 6 months after of any deletions or
changes in test methodologies for any test or examination included in a specialty or
subspecialty, or both, for which the laboratory has been issued a certificate of
accreditation.
Interpretive Guidelines §493.63(a)-(c)
See the section of the SOM beginning at section 6016 for handling changes in ownership,
name, location, personnel and test methodology, or additions or deletions of specialties or
subspecialties that may result in changes in complexity levels for the laboratory.
See the SOM beginning at section 6260 for instructions on handling laboratories that are
going out of business or voluntarily withdrawing from all testing, including laboratories
that are taking these actions while an enforcement action is pending.
Because there is no D-tag assigned to this regulation, you may use D0000 to cite noncompliance with this requirement.
Subpart H--Participation in Proficiency Testing for
Laboratories Performing Nonwaived Testing
Subpart H – General Guidelines
(Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25)
By law, proficiency testing (PT) programs are evaluated initially for CMS approval and
annually thereafter for re-approval. After review, CMS will issue PT program approvals
and/or re-approvals provided they meet the requirements of Subpart I, Proficiency
Testing Programs for Nonwaived Testing. A listing of these programs with the
specialties, subspecialties, and specific analytes for which they are approved is available
on the CMS CLIA web site at
https://www.cms.gov/medicare/quality/clinical-laboratory-
improvement-amendments/proficiency-testing.
A CMS-approved PT program has been evaluated and found to be in compliance with the
requirements of Subpart I and the applicable sections of Subpart H. When a laboratory
experiences problems with PT samples, it resolves them with the PT program. When the
SA experiences problems with an approved program, report all available information to
CMS. CMS renders a decision on the termination or continued approval of the PT
program, as appropriate. The Centers for Disease Control and Prevention may be
requested by CMS to provide technical advice.
History
Rev. 233; Issued: 09-12-25; Effective: 09-12-25; Implementation: 09-12-25
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
5ced1960259bfac5f37790d885ecd8169dc310822c045b136499b7bb70e66740
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.