US · guidance
CMS SOM App. A, Tag A-0537
§482.26(b)(2) - Periodic inspection of equipment must be made and hazards identified must
be properly corrected.
Interpretive Guidelines §482.26(b)(2)
The hospital must have policies and procedures in place to ensure that periodic inspections of
radiology equipment are conducted, current and that problems identified are corrected in a timely
manner. Equipment includes not only devices used to deliver diagnostic or therapeutic
radiologic services, but also exposure meters, badges, or personal radiation monitoring devices
used by staff, as well as equipment the hospital uses to inspect or calibrate devices used to
deliver diagnostic or therapeutic radiologic services. The hospital must ensure that equipment is
inspected in accordance with manufacturer’s instructions and Federal and State laws, regulations,
and guidelines, and hospital policy, as applicable. Inspections and maintenance, including
correction of identified hazards, must be performed by qualified employees (e.g., medical
physicists, qualified biomedical technicians, etc.) or through contractual arrangements with
vendors with appropriate expertise.
Hospitals must follow the manufacturer’s instructions as to how to inspect and maintain
radiologic equipment. This includes acceptance testing (i.e., upon initial installation and after
major upgrades) as well as ongoing inspection and maintenance. Documentation of preventive
maintenance, quality control tests, service records, and major software/hardware upgrades must
be maintained by the hospital and be readily available for inspection.
The hospital must also have a system in place, to identify and remedy equipment hazards in a
timely manner. This system must include, but is not limited to: periodic and consistent
calibration of equipment and, for equipment using ionizing radiation, monitoring of dosimetry
parameters with phantoms to ensure that an accurate dose of radiation is delivered per the
applicable protocol. In addition, hospitals must also have a system to track all modifications
made to the equipment that would significantly impact the accuracy of the dosage delivered.
Any adverse events related to over- or under-dosing must be identified and addressed.
For Information Only - Not Required/Not to be Cited
Requirements for Manufacturers Re: Instructions that Must be Made Available
Below is an FDA summary of its requirements for manufacturers of x-ray systems to make
available to purchasers and, upon request, to other parties, information related to maintenance
of the following types of systems:
• For all diagnostic x-ray systems, manufacturers are required to provide to purchasers, and,
upon request, to others at a cost not to exceed the cost of publication and distribution,
manuals or instruction sheets that include technical and safety information (21 CFR
1020.30(h). This information must include a schedule of the maintenance necessary to
keep the equipment in compliance with §§1020.30, 1020.31, 1020.32, and 1020.33 (21
CFR 1020.30(h)(1)(ii)). Manufacturers are also required to provide to assemblers, and,
upon request, to others at a cost not to exceed the cost of publication and distribution,
instructions for assembly, installation, adjustment, and testing of specified components of
diagnostic x-ray systems adequate to ensure that the products will comply with applicable
provisions of §§1020.30, 1020.31, 1020.32, and 1020.33, when assembled, installed,
adjusted, and tested as directed (21 CFR 1020.30(g)).
• In addition to the requirements applicable to all diagnostic x-ray systems, there are also
other requirements for specific systems:
• Manufacturers of fluoroscopic x-ray systems manufactured on or after June 10, 2006
are required to provide a schedule of maintenance for any system instrumentation
associated with the display of air kerma information necessary to maintain the displays
of air kerma rate and cumulative air kerma within the limits of allowed uncertainty
specified by 21 CFR 1020.32(k)(6). And, if the capability for user calibration of the
display is provided, adequate instructions for such calibration must be supplied (21
CFR 1020.30(h)(6)(i)).
• Manufacturers of computed tomography (CT) systems are required to provide a
specific phantom or phantoms for quality assurance testing of specific system
parameters on these systems (21 CFR 1020.33(d)(1)), and instructions on the use of the
phantom(s), including a schedule of testing appropriate for the system and allowable
variations for the indicated parameters (21 CFR 1020.33(d)(2)).
• Manufacturers of cabinet x-ray systems are required to provide purchasers, and others,
upon request, at a cost not to exceed the cost of preparation and distribution, manuals
and instructions. These documents must include, among other technical and safety
information, a schedule of maintenance necessary to keep the system in compliance
with 21 CFR 1020.40 (21 CFR 1020.40(c)(9)(i)). Cabinet x-ray systems that are
intended to be assembled or installed by the purchaser must be accompanied by
instructions for assembly, installation, adjustment and testing of the cabinet x-ray
adequate to ensure that the system is in compliance with the applicable provisions of
21 CFR 1020.40 when assembled, installed, adjusted and tested as directed (21 CFR
1020.40(c)(9)(ii)).
Survey Procedures §482.26(b)(2)
• Review the inspection records (logs) to verify that periodic inspections are conducted in
accordance with Federal and State laws, regulations and manufacturer’s instructions.
• Verify that inspection and maintenance activities were performed by qualified individuals
• Verify that the maintenance logs show documentation of the calibration upon installation and
after major upgrades or servicing.
• Review with the appropriate personnel the inspection schedule and the mechanism for
identifying hazards, including accurate dosimetry determinations with phantom patients, as
applicable.
• Determine that any problems identified through the testing and maintenance program are
properly corrected in a timely manner and the correction is maintained over time.
History
Rev. 141, Issued: 07-10-15, Effective: 07-10-15, Implementation: 07-10-15
Provenance
- Source
- cms.gov
- Retrieved
- 2026-07-22
- Edition
- som-2026-07-22
- Content hash
b39d21a5a54e36223668e5619d66a53980156928748a26da4d2c777dd65698ea
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.