LCD L40257
Temporary Nontherapeutic Ambulatory Cardiac Monitoring Devices
Coverage Guidance
Title XVIII of the Social Security Act, §1862(a)(1)(A) states that no Medicare payment shall be made for items or services which are not reasonable and necessary for the diagnosis or treatment of illness or injury.
CMS Internet-Only Manual, Pub.100-03, Medicare National Coverage Determinations Manual, Chapter 1, Part 1, §20.15 Electrocardiographic Services CMS Internet-Only Manual Pub. 100-03, National Coverage Determinations, Transmittal 26: Electrocardiographic Services
CMS Internet-Only Manual, Pub. 100-02, Medicare Benefit Policy Manual, Chapter 1, §50 Other Diagnostic or Therapeutic Items or Services
CMS Internet-Only Manual, Pub. 100-02, Medicare Benefit Policy Manual, Chapter 6, §10 Medical and Other Health Services Furnished to Inpatients of Participating Hospitals and §20.3 Encounter Defined
CMS Internet-Only Pub. 100-02, Medicare Benefit Policy Manual, Chapter 15, 60.1 Incident To Physician’s Professional Services, §80 Requirements for Diagnostic X-Ray, Diagnostic Laboratory, and Other Diagnostic Tests and §250 Medical and Other Health Services Furnished to Inpatients of Hospitals and Skilled Nursing Facilities
CMS Internet-Only Manual, Pub. 100-02, Medicate Benefit Policy Manual, Chapter 16, §10 General Exclusions from Coverage
Indications and Limitations of Coverage
Background
Ambulatory electrocardiography provides a record of a patient’s heart rhythm and rate during a specific time frame. The recording could lead to the diagnosis of underlying and undetectable abnormalities. Monitoring allows for the detection of irregular heart rhythms or arrhythmias. These rate disturbances including waveform abnormalities can be missed on a standard electrocardiogram (ECG). The prevalence of cardiac arrythmias in the general population is 1.5% to 5%.1 However, with advancing age and increased incidence of structural heart disease, the frequency, complexity and the prognostic significance of the arrythmias worsens.2 Cardiac arrhythmias are the presenting event in approximately 50% of deaths from cardiovascular disease.1
Temporary nontherapeutic ambulatory cardiac monitoring devices (TNACMD) provide invaluable data regarding arrhythmias, palpitations, syncope and other cardiac monitoring needs. The devices include Holter monitors, event recorders/monitors, patch recorders, external loop recorders (ELR), mobile cardiovascular telemetry (MCT) and mobile cardiovascular outpatient telemetry (MCOT). These distinct devices have different capabilities, varying from intermittent to continuous recordings, and are either patient activated, or rhythm activated recordings. They range in terms of number of leads, time frame length, whether the recordings are being monitored in real time, or if the information recorded needs to be transferred from the device for further interpretation. These devices are not to be utilized in patients at risk for immediate, life-threatening arrythmias (Class I) which require inpatient care.3
The purpose of the LCD is to provide the scope of indications that are supported as reasonable and necessary for the usage of temporary ambulatory nontherapeutic monitoring devices in the appropriate patients.
Coverage
All utilization of electrocardiographic services must align with requirements in NCD 20.15.
Cardiac monitoring is reasonable and necessary when the following criteria are met:
• Temporary (not implanted), AND
• Presence of symptoms suggestive of cardiac arrythmia with symptoms (such as palpitations, dizziness, presyncope, syncope, chest pain or shortness of breath) occurring infrequently (>24 hours between symptomatic episodes),38-40,46,48,54 OR
• Monitoring is necessary to regulate medication management such as antiarrhythmic drug dosage,28,48 OR
• Patients with non-lacunar cryptogenic stroke or stroke or transient ischemic attack (TIA) of undetermined origin to monitor undiagnosed atrial fibrillation (AF) or anticoagulation management,35-37,43,44,50,51,53 OR
• To monitor patients who have had surgical or ablative procedures for arrhythmia,32,47 OR
• To assess for asymptomatic ventricular premature beats or non-sustained ventricular tachycardia in patients with hypertrophic cardiomyopathy, arrhythmogenic right ventricular cardiomyopathy, long QT syndrome, dilated or restrictive cardiomyopathy, congenital heart disease, or Brugada syndrome,42,45,48 OR
• Embolic-appearing pattern of myocardial infarction,41 OR
• Evidence of systemic emboli, OR
• Monitoring for arrhythmic events pre/post-TAVR20-25,55
Device Requirements:
• FDA cleared, AND
• Align with CPT coding standards, FDA clearance, and instructions for their device type including attended surveillance where applicable.
Limitations:
• TNACMD service is medically unnecessary if it offers little or no potential for new clinical data beyond that which has been obtained from a previous test, (e.g., a standard ECG has already established a diagnosis), or if other tests are better suited to obtain clinical data relevant to the patient's condition. The TNACMD should be coordinated with results from standard ECGs, Holter monitor tests, and stress tests.
• The purpose of TNACMD is the long-term monitoring of patients to document suspected or paroxysmal dysrhythmia. Therefore, it is considered medically unnecessary to utilize a TNACMD service when only a standard ECG is required (even if it is used to transmit that ECG to another location).
• The receiving station (for mobile cardiac telemetry) must be staffed on a 24-hour basis with a minimum of an EKG technician or other non-physician staff. Further, such technicians should have immediate, 24-hour access to a physician to review transmitted data and make clinical decisions regarding the patient. The technician should also be instructed as to when and how to contact available facilities to assist the patient in case of emergencies. An answering service/answering machine would not fulfill this requirement. Systems utilizing computers to dial the physician's office, so the physician receives transmission by way of a relay are not covered since there is no 24-hour personnel attendance.
• A test not ordered by a physician or qualified non physician practitioner treating the beneficiary will be denied as not medically necessary.
• It is expected that TNACMD will not be used for the routine daily transmission of ECG rhythm strips, or monitoring, in the absence of identified symptoms necessitating diagnosis as stated in this LCD.
• TNACMD is covered only as diagnostic tests or for evaluating a patient being actively managed on arrhythmic medication.
• TNACMD is not covered for patients in hospitals, emergency rooms, skilled nursing facilities or other specialized facilities, including outpatient or facility-based cardiac monitoring.
• Testing for more than 30 consecutive days is only rarely medically necessary, and the need for the continued testing must be justified by the treating physician. Failure to document arrhythmia during a 30-day test period is not sufficient justification to reimburse a second or subsequent test. It is unlikely to be medically necessary to repeat a second test within a year in the absence of new or recurrent undiagnosed symptoms.
• Event recorders may be patient activated and may not use time-sampling technology. Accordingly, this test will be considered medically unnecessary for any patient who is unresponsive, comatose, severely confused or otherwise unable to recognize symptoms, or activate the recorder (patient activated devices) or unable to participate in the use of the device.
• "Routine" continued monitoring in the absence of treatable symptoms is considered screening and is not medically necessary.
Because the cardiac event detection service requires the diagnosis and evaluation of intermittent arrhythmias, and patients must be continuously attached to pre-symptom loop recorders or be able to be attached at the start of symptoms to post-symptom loop recorders, each patient is required to have a recorder for his/her own exclusive use throughout the duration of the monitoring period. Recorders may not be "shared" amongst 2 or more patients, regardless of the environment or site of the service. Claims for TNACMD will be denied as not medically necessary when patients do not have exclusive use of a recorder for the entire service period (30 days). Cardiac event detection is a packaged service, with device monitoring durations that may be less than or up to 30 days depending on the device used. Tests cannot be billed during any period that overlaps with the billing timeframe of another device covered in this policy, even if the previous test was discontinued due to documented arrhythmias and the patient is reconnected for follow-up therapy or intervention.
Definitions
Temporary Nontherapeutic Ambulatory Cardiac Devices (TNACMD): A portable device which can be attached to the skin to temporarily record the electrical activity of the heart. This allows healthcare providers to monitor the patient over a period of up to 30 days.
Atrial Fibrillation (AF): Supraventricular tachyarrhythmia with uncoordinated atrial electrical activation and ineffectual atrial contraction.
Clinical AF: Symptomatic or asymptomatic AF documented on surface electrocardiogram (ECG).
Subclinical AF or Atrial High-Rate Episode (AHRE): Asymptomatic AF without previous clinical diagnosis of AF.
Cryptogenic Stroke: Subtype of Ischemic stroke without a determined cause or more than one competing cause (30-40% of all ischemic strokes).
Embolic Stroke of Undetermined Source (ESUS): Non-lacunar cryptogenic stroke subtype without identifiable cause despite extensive investigation, including more than 24 hours of cardiac monitoring.
Heart Block: The partial or complete interruption of electrical impulses from the atria to the ventricles (Atrioventricular (AV) block), where the rate is slower, or the rhythm is abnormal.
Implantable Cardioverter Defibrillator (ICD): A device implanted in the chest to detect and treat life-threatening cardiac arrhythmias by delivering electrical shocks or pacing.
Intraventricular Conduction Delay (IVCD): Prolongation of the QRS exceeding the typical upper limit of normal but does not meet criteria for a specific conduction block.
Junctional Rhythm: Abnormal rhythm originating from the AV node or the Bundle of HIS (HB).
Non-lacunar Stroke: Cerebrovascular accident presumed to be caused by a thromboembolic event from the heart or the large extracranial arteries.
Acute Coronary Syndrome (ACS): Acute myocardial infarction or ischemia.
Silent Myocardial Ischemia: Asymptomatic cardiac ischemia on testing without angina or an anginal equivalent.
Systemic Emboli: Embolic events affecting organs or tissues outside the heart, confirmed by clinical, imaging, or laboratory findings.
Summary of Evidence
Literature Analysis
This summary of the evidence has been formatted by the type of comparator, outcome category, (e.g., diagnostic yield, undesirable effects), and study design. The literature analysis emphasizes the research designs representative of the body of evidence (systematic reviews, meta-analyses) and the primary studies most applicable to specific outcomes e.g., randomized controlled trials (RCTs) for diagnostic yield. Multiple publications from the same dataset were grouped together in the analysis regardless of study design. The GRADE (Grading of Recommendations, Assessment, Development, and Evaluation) approach domains of study limitations (risk of bias), indirectness (applicability), imprecision, inconsistency, and publication bias formed the basis of appraisal of the certainty of evidence [see Attachments, Tables 1 and 2].
Description of Studies
This description of studies is supplemented by tables 3–8 (see Attachments) that further summarize the characteristics, findings, and assessments of included studies.
This evidentiary review included 12 systematic reviews, most with meta-analysis [Table 3]. These reviews synthesized studies for all types of TNACMD. Only 2 reviews were limited to the analysis of RCTs. The number of primary studies and participants in the reviews ranged from 3 to 50 and 1149 to 135,300, respectively. The patient populations varied from asymptomatic older adults to individuals diagnosed with stroke or TIA. The detection of AF was a common primary outcome [Table 4]. Implantable cardiac monitors were the most frequently employed comparator. Five systematic reviews assessed clinical outcomes. Three syntheses reported on undesirable effects. Two reviews described the effect of TNACMD on clinical management. Five of the systematic reviews provided data about potential effect modifiers, patient selection, or the timing of TNACMD.
A total of 9 RCTs were included in the evidence review [Table 5]. Most trials were multi-centered and took place in North America. The number of participants varied widely from 21 to 11,931. The mean/median ages of participants ranged from 64.1 to 75 years. These studies included multi-morbid patients without known AF who had cryptogenic ischemic stroke, TIA, risk factors for stroke following cardiac surgery, or received opportunistic screening. Five trials assessed long-term cardiac monitors (LTCM) i.e., patch devices. Three RCTs evaluated Holter type monitoring devices. Four studies investigated event monitors e.g., ELR. There were no RCTs concerning MCT or MCOT devices that met eligibility criteria. Comparator interventions included usual care, implantable cardiac monitors, and between different types of TNACMD. Most studies reported diagnostic yield (detection of AF) as a primary endpoint [Table 6]. Primary and secondary outcomes were assessed at follow-up time points between 14 days and 15 months. A single large retrospective cohort study of Medicare beneficiaries (N = 287,789) was also included in the analysis.
There were 6 non-randomized studies of an intervention (NRSI) identified that investigated outcomes associated with the use of MCT devices on single cohorts that received TAVR [Table 7]. Four studies were conducted prospectively. Two trials evaluated the impact of TNACMD when used during the peri-procedural period (i.e., before and after TAVR). A single study assessed the value of a 1-week monitoring period prior to the implantation of a TAVR device. Three case series reported on the results of the extended application of real-time MCT ECG monitoring immediately following TAVR implantation. Five of the trials were performed at single centers in the USA or Canada. A fifth study involved 2 locations. Multiple types of MCT devices were used in the studies, with BodyGuardian® and Pocket-ECG the most frequently reported. Monitoring periods were predominantly in the range of 28-30 days. Most studies reported on the incidence and type of arrhythmic events and consequential therapeutic changes [Table 8]. Other less frequently reported outcomes included feasibility (compliance, notifications) and mortality. All the studies were small, ranging from 23-192 participants obtained from convenience samples.
Effects of TNACMD
TNACMD compared to Standard of Care (SOC)
Diagnostic Yield
Two evidence syntheses analyzed data from primary studies that compared various TNACMD with SOC including short-term cardiac monitoring.4,5 Dahal et al.4 performed a meta-analysis of 4 RCTs (N = 1149) that evaluated the prolonged monitoring (LTCM, MCOT, ELR) ≥7 days in patients with cryptogenic stroke or TIA. The reviewers found prolonged cardiac monitoring of ≥7 days compared to shorter cardiac monitoring of ≤48 hours duration increased the detection of AF (≥30 seconds duration) in patients after cryptogenic stroke or TIA (13.8% vs. 2.5%; odds ratio [OR] 6.4; 95% confidence interval [CI], 3.50, 11.73; P < 0.00001; I2 = 0%]. It also increased the odds of AF detection of any duration (22.6% vs. 5.2%; 5.68 [3.3, 9.77]; P < 0.00001; I2 = 0%). This meta-analysis has several limitations including lack of patient-level data, different duration of follow-up among studies, and variations in type of device. It was not possible to ascertain which device and follow-up duration worked better compared to the others.
A health technology assessment (HTA) employed a network meta-analysis, with SOC (short-term Holter monitoring) as the common comparator, in assessing the relative effectiveness of LTCM compared with ELRs.5 This HTA included 12 non-randomized studies of interventions (NRSI) that reported on patients with palpitations, syncope, dizziness, stable chronic heart failure, ischemic and non-ischemic stroke, cryptogenic stroke, TIA, or known/suspected dysrhythmias. Both types of devices were more effective than SOC. There was no substantial difference between them in their ability to detect symptoms (risk difference 0.01; 95% CI: –0.18, 0.20). Using GRADE for network meta-analysis, the quality of the evidence was evaluated as low. Additional constraints were that all primary studies were observational designs, and only indirect comparisons could be obtained.
Six RCTs reported on the detection of AF by different TNACMD as compared to SOC that included 24-hour Holter monitoring.6-11 These studies included patients without known AF who had cryptogenic ischemic stroke, TIA, risk factors for stroke following cardiac surgery, or received asymptomatic screenings. Monitoring periods ranged from 14 to 90 days, with detection endpoints extending to as much as 15.3 months. The TNACMD strategies were superior in detecting AF versus short-term monitoring at all follow-up periods (TNACMD = median 16.3% [14-19.6], SOC = 2.1% [1.7-5]) except for opportunistic screening (TNACMD = 0.7%, SOC = 0.6%).
Cumulative AF or atrial flutter (AFL) was calculated in 3 RCTs. Within the first 30 days of monitoring, AF/AFL of >6 and >24 hours durations were detected in 8.6% and 3.1% of the TNACMD group versus 0% in the SOC group, respectively.7 Paroxysmal atrial fibrillation (PAF) of any duration was identified in 1 RCT at 28 days (14% in the TNACMD group and 2.1% in the SOC group)9 and through 90 days (44% in the TNACMD group and 4% in the SOC group).8
Clinical Management
Dahal et al.4 meta-analyzed data from 4 RCTs (N = 1149). Patients diagnosed with cryptogenic stroke or TIA who underwent prolonged monitoring (LTCM, ELR, MCOT) were more likely to be on anticoagulation at follow-up than those receiving SOC (2.21[1.52, 3.21]; P < 0.0001; I2, 0%).
There were 4 RCTs that explored the impact of TNACMD on the initiation of oral anti-coagulant (OAC) therapy.6,8-10 Higgins et al.8 found absolute differences favoring the TNACMD group of 16% at 14- and 90-day time points. Another RCT reported an absolute difference of 7.5% in favor of TNACMD compared to SOC at 90 days.6 Two RCTs provided insufficient details (uncertainties about the absolute effects and time frame) to permit conclusions about the effects of TNACMD on clinical management.9,10
Mortality, Stroke, Major Adverse Cardiac Events
A single systematic review and meta-analysis of 4 RCTs involving 1149 patients with cryptogenic stroke or TIA who underwent prolonged monitoring (LTCM, ELR, MCOT) or SOC (short-term Holter monitoring) found no differences in mortality (1.33 [0.29, 6.00]; P = 0.71; I2 = 0%] or recurrent stroke or TIA (0.78 [0.40, 1.55]; P = 0.48; I2, 0%).4
Two trials investigated mortality, finding no significant difference with SOC at 90 days between patients assigned to a 7-day ELR8 or 14-day patch monitor.9
There was no evidence that screening for AF using a 14-day continuous TNACMD reduces recurrent stroke or TIA.8,9 The data were from 2 small RCTs (N = 100, 116), and event rates were low. This may have resulted in underestimating the effect of TNACMD.
Healthcare Utilization
There was no indication, from a single RCT, that screening for AF using a 14-day continuous TNACMD in people ≥70 years of age seen in primary care practice reduces stroke-related hospitalizations.10 However, the event rates were low, which may have resulted in underestimating the effect of TNACMD.
Undesired Effects
There were no serious adverse events associated with the use of the ELR and patch devices.7,8,10 A single RCT reported 10.4% of participants experienced a device-related adverse event due to skin irritation from the adhesive material of the wearable patch within 30 days of randomization.7
Adherence
Data from a single trial found that among the patients in the intervention group who started to undergo monitoring, 233 of 284 (82.0%) completed 3 or more weeks of monitoring.6
Effect Modifiers
No studies were identified.
TNACMD compared to Implantable Cardiac Monitors
Diagnostic Yield
Six systematic reviews evaluated the effectiveness of implantable loop recorder (ILR) versus TNACMD for identifying AF in patients with any type of stroke (e.g., cryptogenic, ischemic) or TIA.12-17 Broadly, these reviews concluded that for patients with sufficient cognitive and physical ability to comply with the use of TNACMD, a 1-month duration can capture a significant proportion of AF and should be considered in place of ILRs. The choice of specific type of TNACMD should be made based on availability, patient’s preferences, and clinical judgment. ILRs can be considered in patients where a prolonged duration of monitoring (>30 days) is anticipated, if a TNACMD fails to detect any AF after 4 weeks of monitoring, or if there are anticipated issues with compliance.
A single RCT sought to determine, in patients with a recent ischemic stroke, whether 12 months of ILR monitoring detected more occurrences of AF compared with conventional ELR monitoring for 30 days.18 There was a statistically significant between-group difference in the rate of detection of AF or flutter lasting >2 minutes by 12 months (15.3% in the ILR group vs 4.7% in the prolonged ELR group).
Clinical Management
Al Qurashi et al.13 conducted a systematic review and meta-analysis to assess the efficacy of implantable cardiac monitors (ICMs) compared to a range of TNACMD in detecting post-stroke AF and the subsequent use of OAC therapy. The review included 3 RCTs of 1233 patients with any type of stroke and no prior AF or flutter. The ICM arm had significantly higher usage of oral anticoagulants as compared to the TNACMD arm. (RR = 2.76, 95% CI: 1.89, 4.02, P < 0.05). Limitations of this review included the small number of eligible studies and the need to assume the baseline characteristics of patients in all the included trials were similar.
Mortality, Stroke, Major Adverse Cardiac Events
A single systematic review and meta-analysis reported showed there were no significant differences between ICMs and TNACMD in the reduction of mortality (RR = 0.67, 95% CI: 0.31, 1.45; P = 0.42; I2 = 0%) and in the recurrence of stroke (RR = 0.73, 95% CI: 0.48, 1.10; P = 0.14; I2 = 0%).13
Healthcare Utilization
No studies found.
Undesired Effects
A systematic review with meta-analysis found ICM usage was associated with a higher incidence of mild to moderate adverse events (RR = 10.52, 95% CI:1.35, 82.14; P = 0.02) and a higher number of severe adverse events as compared to the use of TNACMD (RR = 7.61, 95% CI: 1.36, 42.51; P = 0.02).13
Effect Modifiers
Noubiap et al.17 performed a systematic review and meta-analysis that explored factors influencing AF detection rates. The reviewers found 1) a steady increase of AF rates with duration of monitoring; 2) higher rates of AF in patients with ESUS compared to those with cryptogenic stroke; and 3) the association of older age, CHA2DS2-VASc score, PA-TDI Interval, left atrial enlargement, P wave maximal duration, prolonged PR interval and atrial runs with higher rates of AF detection.
Comparison of Different Types of TNACMD
A network meta-analysis found that LTCM and ELR devices were equally effective in their ability to detect symptomatic cardiac arrhythmias.5
Reynolds, et al.19 retrospectively analyzed the USA National Medicare claims data of 287,789 beneficiaries without a preceding established arrhythmia diagnosis. The aim of the study was to make comparisons among different device types (ELRs, LTCM, Holter, and MCT) across a range of outcomes. The device-specific analysis showed that compared to Holter, ELR, MCT, or other LTCM manufacturers, a specific LTCM (Zio® XT 14-day patch, iRhythm Technologies, San Francisco, CA) had the highest adjusted odds of AF detection and lowest adjusted odds of TNACMD retesting. Findings were consistent for specific arrhythmia diagnoses of ventricular tachycardia, AV block, and PAF. The annualized all-cause inpatient hospitalizations during the follow-up period were lowest in the LTCM and Holter cohorts (mean = 0.45 stays for both cohorts), followed by the ELR and MCT cohorts (mean 0.60 stays for both cohorts). Annualized follow-ups for all-cause ED visits were lowest in the LTCM cohort (mean = 0.70 visits). Meanwhile, outpatient visits were lowest in the Holter and ELR cohorts (mean = 24.5 visits for both cohorts); however, patients in the LTCM cohort experienced the smallest increase in outpatient visits over baseline (mean change = 3.11 for the LTCM cohort). This study included limitations inherent in retrospective observational designs, the inability to adjust for the duration of monitoring, potential for industry funded bias (6 of 8 authors were affiliated with the sponsor), and the likelihood of confounding due to the influence of patient, provider, and system factors.
Non-Comparative Studies
Diagnostic Yield
Six NRSI reported on the incidence, duration, and type (e.g., AF, bradycardia, high-degree AV block) of arrhythmic events identified by TNACMD either before and/or after implantation of a TAVR device.20-25 One methodologically weak study found arrhythmic events were diagnosed in almost half of patients.20 The other 5 studies reported on a more narrow range (7.3% – 10%) of new arrhythmic events.
Clinical Management
Three NRSI described changes in therapeutic management – mostly permanent pacemaker implantation – as a result of extended cardiac monitoring in patients undergoing TAVR.20,21,23 The implementation of TNACMD prior to TAVR resulted in a change in therapeutic management (anti-coagulation medication, permanent pacemaker) for more than a quarter of patients.20 The use of TNACMD post-TAVR resulted in smaller proportions (3.1% – 6%) of patients undergoing permanent pacemaker implantation.21,23
Mortality, Stroke, Major Adverse Cardiac Events
A small non-comparative study (N=54) did not record any mortalities for patients using an MCT monitor for 30 days following TAVR.23
Undesired Effects
No studies were identified.
Adherence
Two NRSI assessed patient compliance with using MCT technologies prior to or immediately following the TAVR procedure.21,23 Overall adherence for at least 14 days was high (87.5% - 96%).
Effect Modifiers
No studies were identified.
Summary of Main Results
Prolonged cardiac monitoring with TNACMD is more effective than SOC in the detection of AF in patients after stroke or TIA, or with symptoms associated with an increased risk of AF. For patients with the ability to comply with the use of TNACMD, a 30-day duration can capture a significant proportion of AF and should be considered in lieu of ILRs. Patients using TNACMD are more likely to receive guideline-concordant anti-coagulant treatment compared to those undergoing SOC. In contrast, individuals with ICMs are more likely to be prescribed timely anti-coagulant therapy than those wearing TNACMD. There were no significant differences between SOC, ICMs and TNACMD in the reduction of mortality, or the recurrence of stroke or TIA. The likelihood of AF detection steadily increases with the duration of monitoring and patient age. It is uncertain whether any type of TNACMD is clinically superior to other externally wearable devices. TNACMD may improve the detection of conduction disturbances pre/post-TAVR, which may then result in meaningful changes in therapeutic management.
Applicability of Evidence
The included studies were highly applicable to the Medicare beneficiary population. The largest study explicitly focused on the Medicare population. Most trials were multi-centered and took place in North America. The mean/median ages of participants ranged from 64.1 to 84 years. The included studies reported on multi-morbid patients.
Quality of Evidence
The overall certainty of evidence was rated as moderate for clinical, healthcare management/utilization, and undesirable effects outcomes. There were some concerns about the risk of bias (selection and performance bias). There was no serious indirectness, inconsistency, imprecision, or risk of publication bias.
The certainty of evidence for the comparative effectiveness of different TNACMD was judged to be very low (observational data, high risk of confounding, and the potential for industry funding bias).
The certainty of evidence for the use of TNACMD for patients undergoing TAVR was low to very-low (non-comparative observational studies, small sample sizes, high risk of selection bias, and limited generalizability).
Clinical Guidelines and Positions of National and Specialty Organizations
A total of 22 publications developed by 15 professional societies or agencies were identified to supplement the summary of evidence [see Attachments, Table 9].34-55 These guidelines encompassed a broad range of disorders (stroke, TIA, cardiac arrhythmias, cardiac conduction delay, syncope, silent cerebrovascular disease, hypertrophic cardiomyopathy, severe aortic stenosis) and monitoring during AF ablation.
Collectively, there was a strong consensus in favor of extended cardiac monitoring in situations where AF cannot readily be diagnosed with standard ECG. There was limited explicit guidance regarding the optimal monitoring devices. Most guidance recommended that the frequency of symptoms should dictate the type of monitoring device. A single guideline proposed a clinical workflow commencing with continuous (short-term 24 hour and up to 7 days) ambulatory ECG monitoring, which if unsuccessful, is followed by intermittent external loop recording (long term from weeks to months). For those individuals remaining undiagnosed after prolonged noninvasive monitoring, ILR may be necessary.48 An American College of Cardiology expert panel favored prolonged ECG monitoring (>14 days) during the post-procedural period following TAVR due to the increased risk of arrhythmic events.55
Analysis of Evidence
This analysis of evidence provides moderate certainty evidence that TNACMD demonstrate consistent improvements in diagnostic yield and clinical management without significant undesirable effects in patients who are representative of U.S. Medicare beneficiaries. TNACMD are as effective in detecting arrhythmias as ICM and have increased the odds of AF detection as compared to monitoring for <48 hours. There is consistent evidence (low and very-low certainty) that the application of TNACMD during the peri-procedural period for patients undergoing TAVR may meaningfully improve care management for up to 25% of patients. Professional guidelines include evidence-informed recommendations for the appropriate timing and the type of monitoring device in clinical practice. Consequently, this evidentiary review advises for the inclusion of TNACMD as reasonable and necessary for the detection of cardiac arrhythmias in appropriately selected patients.
Associated Information
N/A
Bibliography
• Desai DS, Hajouli S. Arrhythmias. In: StatPearls [Internet]. StatPearls Publishing; 2024. Updated June 5, 2023. Accessed April 13, 2026.
• Mirza M, Strunets A, Shen WK, Jahangir A. Mechanisms of arrhythmias and conduction disorders in older adults. Clin Geriatr Med. 2012;28(4):555-573. doi:10.1016/j.cger.2012.08.005
• Drew BJ, Califf RM, Funk M, et al. AHA scientific statement: Practice standards for electrocardiographic monitoring in hospital settings. J Cardiovasc Nurs. 2005;20(2):76-106. doi:10.1097/00005082-200503000-00003
• Dahal K, Chapagain B, Maharjan R, et al. Prolonged cardiac monitoring to detect atrial fibrillation after cryptogenic stroke or transient ischemic attack: A meta-analysis of randomized controlled trials. Ann Noninvasive Electrocardiol. 2016;21(4):382-388. doi:10.1111/anec.12319
• Health Quality Ontario. Long-term continuous ambulatory ECG monitors and external cardiac loop recorders for cardiac arrhythmia: a health technology assessment. Ont Health Technol Assess Ser. 2017;17(1):1-56.
• Gladstone DJ, Spring M, Dorian P, et al. Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014;370(26):2467-2477. doi:10.1056/NEJMoa1311376
• Ha ACT, Verma S, Mazer CD, et al. Effect of continuous electrocardiogram monitoring on detection of undiagnosed atrial fibrillation after hospitalization for cardiac surgery: A randomized clinical trial. JAMA Netw Open. 2021;4(8):e2121867. doi:10.1001/jamanetworkopen.2021.21867
• Higgins P, MacFarlane PW, Dawson J, et al. Noninvasive cardiac event monitoring to detect atrial fibrillation after ischemic stroke: a randomized, controlled trial. Stroke. 2013;44(9):2525-2531. doi:10.1161/STROKEAHA.113.001927
• Kaura A, Sztriha L, Chan FK, et al. Early prolonged ambulatory cardiac monitoring in stroke (EPACS): An open-label randomised controlled trial. Eur J Med Res. 2019;24(1):25. doi:10.1186/s40001-019-0383-8
• Singer DE, Atlas SJ, Go AS, et al. ReducinG stroke by screening for Undiagnosed atrial fibrillation in elderly inDividuals (GUARD-AF): Rationale and design of the GUARD-AF randomized trial. Am Heart J. 2022;249:76-85. doi:10.1016/j.ahj.2022.04.005
• Wachter R, Gröschel K, Gelbrich G, et al. Holter-electrocardiogram-monitoring in patients with acute ischaemic stroke (Find-AFRANDOMISED): an open-label randomised controlled trial [published correction appears in Lancet Neurol. 2017 Apr;16(4):261. doi: 10.1016/S1474-4422(17)30036-4.]. Lancet Neurol. 2017;16(4):282-290. doi:10.1016/S1474-4422(17)30002-9
• Afzal MR, Gunda S, Waheed S, et al. Role of outpatient cardiac rhythm monitoring in cryptogenic stroke: A systematic review and meta-analysis. Pacing Clin Electrophysiol. 2015;38(10):1236-1245. doi:10.1111/pace.12688
• Al Qurashi AA, Rasheed F, Siddiqi AR, et al. Insertable cardiac monitoring devices versus external cardiac monitoring for detecting atrial fibrillation and preventing the recurrence of stroke: A systematic review and meta-analysis. J Electrocardiol. 2023;77:29-36. doi:10.1016/j.jelectrocard.2022.11.007
• Dussault C, Toeg H, Nathan M, et al. Electrocardiographic monitoring for detecting atrial fibrillation after ischemic stroke or transient ischemic attack: systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2015;8(2):263-269. doi:10.1161/CIRCEP.114.002521
• Jiang H, Tan SY, Wang JK, et al. A meta-analysis of extended ECG monitoring in detection of atrial fibrillation in patients with cryptogenic stroke. Open Heart. 2022;9(2):e002081. doi:10.1136/openhrt-2022-002081
• Korompoki E, Del Giudice A, Hillmann S, et al. Cardiac monitoring for detection of atrial fibrillation after TIA: A systematic review and meta-analysis. Int J Stroke. 2017;12(1):33-45. doi:10.1177/1747493016669885
• Noubiap JJ, Agbaedeng TA, Kamtchum-Tatuene J, et al. Rhythm monitoring strategies for atrial fibrillation detection in patients with cryptogenic stroke: A systematic review and meta-analysis. Int J Cardiol Heart Vasc. 2021;34:100780. doi:10.1016/j.ijcha.2021.100780
• Buck BH, Hill MD, Quinn FR, et al. Effect of implantable vs prolonged external electrocardiographic monitoring on atrial fibrillation detection in patients with ischemic stroke: the PER DIEM randomized clinical trial. JAMA. 2021;325(21):2160-2168. doi:10.1001/jama.2021.6128
• Reynolds MR, Passman R, Swindle J, et al. Comparative effectiveness and healthcare utilization for ambulatory cardiac monitoring strategies in Medicare beneficiaries. Am Heart J. 2024;269:25-34. doi:10.1016/j.ahj.2023.12.002
• Asmarats L, Nault I, Ferreira-Neto AN, et al. Prolonged continuous electrocardiographic monitoring prior to transcatheter aortic valve replacement: the PARE study. JACC Cardiovasc Interv. 2020;13(15):1763-1773. doi:10.1016/j.jcin.2020.03.031
• Natarajan MK, Sheth TN, Wijeysundera HC, et al. Remote ECG monitoring to reduce complications following transcatheter aortic valve implantations: the Redirect TAVI study. Europace. 2022;24(9):1475-1483. doi:10.1093/europace/euac042
• Ream K, Sandhu A, Valle J, et al. Ambulatory rhythm monitoring to detect late high-grade atrioventricular block following transcatheter aortic valve replacement. J Am Coll Cardiol. 2019;73(20):2538-2547. doi:10.1016/j.jacc.2019.02.068
• Skaf M, Makki N, Coressel A, et al. Rhythm disturbances after transcatheter aortic valve replacement: A strategy of surveillance. Cardiovasc Revasc Med. 2020;21(4):475-478. doi:10.1016/j.carrev.2019.07.017
• Tian Y, Padmanabhan D, McLeod CJ, et al. Utility of 30-day continuous ambulatory monitoring to identify patients with delayed occurrence of atrioventricular block after transcatheter aortic valve replacement. Circ Cardiovasc Interv. 2019;12(12):e007635. doi:10.1161/CIRCINTERVENTIONS.118.007635
• Winter JL, Healey JS, Sheth TN, et al. Remote ambulatory cardiac monitoring before and after transcatheter aortic valve replacement. CJC Open. 2020;2(5):416-419. doi:10.1016/j.cjco.2020.04.006
• Ho JS, Ho ES, Yeo LL, et al. Use of wearable technology in cardiac monitoring after cryptogenic stroke or embolic stroke of undetermined source: A systematic review. Singapore Med J. 2024;65(7):370-379. doi:10.4103/singaporemedj.SMJ-2022-143
• Jonas DE, Kahwati LC, Yun JDY, et al. Screening for atrial fibrillation with electrocardiography: Evidence report and systematic review for the US Preventive Services Task Force. JAMA. 2018;320(5):485-498. doi:10.1001/jama.2018.4190
• Kahwati L, Asher GN, Kadro ZO, et al. Screening for Atrial Fibrillation: An Evidence Review for the U.S. Preventive Services Task Force. Agency for Healthcare Research and Quality (US); 2022.
• Sposato LA, Cipriano LE, Saposnik G, et al. Diagnosis of atrial fibrillation after stroke and transient ischaemic attack: A systematic review and meta-analysis. Lancet Neurol. 2015;14(4):377-387. doi:10.1016/S1474-4422(15)70027-X
• Eysenck W, Freemantle N, Sulke N. A randomized trial evaluating the accuracy of AF detection by four external ambulatory ECG monitors compared to permanent pacemaker AF detection. J Interv Card Electrophysiol. 2020;57(3):361-369. doi:10.1007/s10840-019-00515-0
• Lopes RD, Atlas SJ, Go AS, et al. Effect of screening for undiagnosed atrial fibrillation on stroke prevention. J Am Coll Cardiol. 2024;84(21):2073-2084. doi:10.1016/j.jacc.2024.08.019
• Steinhubl SR, Waalen J, Edwards AM, et al. Effect of a home-based wearable continuous ECG monitoring patch on detection of undiagnosed atrial fibrillation: The mSToPS randomized clinical trial. JAMA. 2018;320(2):146-155. doi:10.1001/jama.2018.8102
• Wachter R, Gröschel K, Gelbrich G, et al. Holter-electrocardiogram-monitoring in patients with acute ischaemic stroke (Find-AFRANDOMISED): An open-label randomised controlled trial. Lancet Neurol. 2017;16(4):282-290. doi:10.1016/S1474-4422(17)30002-9
• Culebras A, Messé SR, Chaturvedi S, Kase CS, Gronseth G. Summary of evidence-based guideline update: prevention of stroke in nonvalvular atrial fibrillation: Report of the Guideline Development Subcommittee of the American Academy of Neurology. Neurology. 2014;82(16):1481. doi:10.1212/wnl.0000000000000465
• Culebras A, Messe SR, Chaturvedi S, et al. Summary of evidence-based guideline update: Prevention of stroke in nonvalvular atrial fibrillation: report of the Guideline Development Subcommittee of the American Academy of Neurology. Neurology. 2014;82(8):716-724. doi:10.1212/WNL.0000000000000145
• Jauch EC, Saver JL, Adams HP Jr, et al. Guidelines for the early management of patients with acute ischemic stroke: A guideline for healthcare professionals from the American Heart Association/American Stroke Association. Stroke. 2013;44(3):870-947. doi:10.1161/STR.0b013e318284056a
• January CT, Wann LS, Alpert JS, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2014;64(21):e1-e76. doi:10.1016/j.jacc.2014.03.022
• Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm. 2018;15(10):e73-e189. doi:10.1016/j.hrthm.2017.10.036
• Shen WK, Sheldon RS, Benditt DG, et al. 2017 ACC/AHA/HRS guideline for the evaluation and management of patients with syncope: Executive summary: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2017;70(5):620-663. doi:10.1016/j.jacc.2017.03.002
• Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2019;140(8):e382-e482. doi:10.1161/CIR.0000000000000628
• Smith EE, Saposnik G, Biessels GJ, et al. Prevention of stroke in patients with silent cerebrovascular disease: a scientific statement for healthcare professionals from the American Heart Association/American Stroke Association. Stroke. 2017;48(2):e44-e71. doi:10.1161/STR.0000000000000116
• Ommen SR, Mital S, Burke MA, et al. 2020 AHA/ACC guideline for the diagnosis and treatment of patients with hypertrophic cardiomyopathy: Executive summary: A report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2020;142(25):e533-e557. doi:10.1161/CIR.0000000000000938
• Kleindorfer DO, Towfighi A, Chaturvedi S, et al. 2021 guideline for the prevention of stroke in patients with stroke and transient ischemic attack: A guideline from the American Heart Association/American Stroke Association. Stroke. 2021;52(7):e364-e467. doi:10.1161/STR.0000000000000375
• Joglar JA, Chung MK, Armbruster AL, et al. 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation: A report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2024;149(1):e1-e156. doi:10.1161/CIR.0000000000001193
• Gersh BJ, Maron BJ, Bonow RO, et al. 2011 ACCF/AHA guideline for the diagnosis and treatment of hypertrophic cardiomyopathy: A report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2011;58(25):e212-e260. doi:10.1016/j.jacc.2011.06.011
• Pedersen CT, Kay GN, Kalman J, et al. EHRA/HRS/APHRS expert consensus on ventricular arrhythmias. Europace. 2014;16(9):1257-1283. doi:10.1093/europace/euu194
• Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14(10):e275-e444. doi:10.1016/j.hrthm.2017.05.012
• Steinberg JS, Varma N, Cygankiewicz I, et al. 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Ann Noninvasive Electrocardiol. 2017;22(3):e12447. doi:10.1111/anec.12447
• Brignole M, Moya A, de Lange FJ, et al. 2018 ESC guidelines for the diagnosis and management of syncope. Eur Heart J. 2018;39(21):1883-1948. doi:10.1093/eurheartj/ehy037
• Hindricks G, Potpara T, Dagres N, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. 2021;42(5):373-498. doi:10.1093/eurheartj/ehaa612
• Andrade JG, Aguilar M, Atzema C, et al. The 2020 Canadian Cardiovascular Society/Canadian Heart Rhythm Society comprehensive guidelines for the management of atrial fibrillation. Can J Cardiol. 2020;36(12):1847-1948. doi:10.1016/j.cjca.2020.09.001
• Davidson KW, Barry MJ, Mangione CM, et al. Screening for atrial fibrillation: US Preventive Services Task Force recommendation statement. JAMA. 2022;327(4):360-367. doi:10.1001/jama.2021.23732
• Rubiera M, Aires A, Antonenko K, et al. European Stroke Organisation (ESO) guideline on screening for subclinical atrial fibrillation after stroke or transient ischaemic attack of undetermined origin. Eur Stroke J. 2022;7(3):VI. doi:10.1177/23969873221099478
• National Institute for Health and Care Excellence. Atrial Fibrillation: Diagnosis and Management. NICE Guideline [NG196]; 2021. Accessed April 13, 2026.
• Rodés-Cabau J, Ellenbogen KA, Krahn AD, et al. Management of conduction disturbances associated with transcatheter aortic valve replacement: JACC Scientific Expert Panel. J Am Coll Cardiol. 2019;74(8):1086-1106. doi:10.1016/j.jacc.2019.07.014
History
Version 7
Provenance
- Source
- cms.gov
- Retrieved
- 2026-08-26
- Edition
- mcd-2026-08-26
- Content hash
1f4f21eaf0e8d9050339dfce6ac14db343f31f9dd67e98001c0982b64cd87ce2
The link goes to the issuing authority’s own document — the one we read to produce this record. Where a source publishes whole titles rather than sections, your browser may need a moment to jump to the provision.
Unofficial copy of government-published law, reproduced from official sources with full provenance. Not an official publication; verify against official sources before relying on it in a filing. Records in the 'guidance' corpus, and only that corpus, are sub-regulatory (interpretive guidelines, survey procedures) and are not binding law. Validity bounds follow each jurisdiction's declared temporalBasis.